Suppr超能文献

哈达玛变换在胶束电动毛细管色谱中的应用。

Application of Hadamard transformation to MEKC.

作者信息

Hata Kazuki, Kaneta Takashi, Imasaka Totaro

机构信息

Department of Applied Chemistry, Graduate School of Engineering, Kyushu University, Motooka, Fukuoka, Japan.

出版信息

Electrophoresis. 2007 Feb;28(3):328-34. doi: 10.1002/elps.200600183.

Abstract

The Hadamard transform (HT) technique, which permits the S/N in CE to be improved, was applied to MEKC. Multiple sample injection of fluorescent analytes according to a Hadamard code sequence was performed using an optically gated sample injection technique, in which a sample plug was produced based on photodegradation by irradiation with an intense laser beam. The capillary and reservoirs were filled with a sample solution containing buffer components and SDS as a pseudostationary phase. A preliminary study confirmed that fluorescein ion could be photobleached in the presence of SDS. The optically gated sample injection technique was then applied to multiple sample injection, based on a Hadamard matrix. The S/N in the electropherogram obtained by HT-MEKC was improved substantially compared to that obtained by a single injection method. When the technique was applied to the separation of several amino acids labeled with FITC, the S/N ratio for each amino acid was enhanced, without any evidence of degradation in separation resolution. Moreover, HT-MEKC was applied to the analysis of amino acids contained in a Japanese beverage, resulting in improved S/Ns for the amino acids.

摘要

允许提高毛细管电泳信噪比的哈达玛变换(HT)技术被应用于胶束电动毛细管色谱(MEKC)。使用光控进样技术根据哈达玛编码序列对荧光分析物进行多次进样,其中通过用强激光束照射基于光降解产生样品塞。毛细管和储液器中充满含有缓冲成分和作为假固定相的十二烷基硫酸钠(SDS)的样品溶液。一项初步研究证实,在SDS存在下荧光素离子可被光漂白。然后基于哈达玛矩阵将光控进样技术应用于多次进样。与单次进样方法相比,通过HT-MEKC获得的电泳图中的信噪比有了显著提高。当该技术应用于分离几种用异硫氰酸荧光素(FITC)标记的氨基酸时,每种氨基酸的信噪比均得到提高,且分离分辨率没有任何下降的迹象。此外,HT-MEKC被应用于分析一种日本饮料中所含的氨基酸,从而提高了这些氨基酸的信噪比。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验