The effects of abutment wall height, platform size, and screw access channel filling method on resistance to dislodgement of cement-retained, implant-supported restorations.

作者信息

Emms Mark, Tredwin Christopher J, Setchell Derrick J, Moles David R

机构信息

Department of Conservative Dentistry, Eastman Dental Institute, University College London, London, UK.

出版信息

J Prosthodont. 2007 Jan-Feb;16(1):3-9. doi: 10.1111/j.1532-849X.2006.00150.x.

Abstract

PURPOSE

Factors affecting the retention of fixed prostheses to natural abutments are well understood. In contrast, little is known concerning factors influencing the retention of fixed prostheses cemented to implant abutments. The purpose of this study was to investigate the effect that varying implant abutment wall height, platform size, and screw access channel filling method has on the retention of castings cemented to implant abutments using TempBond.

MATERIALS AND METHODS

Four 15 degrees preangled abutments (Nobel Biocare Replace Select Esthetic) of each platform size--narrow (NP), regular (RP), and wide (WP)--were used. In each group of abutments the screw access axial wall was either unadjusted, one-third removed, two-thirds removed, or completely removed. The screw access channels were either fully or partially filled with Memosil, a vinyl polysiloxane impression material. For each abutment a casting was constructed that incorporated an attachment to allow removal. Castings were cemented to abutments with TempBond. The tensile force required to separate the cemented castings from the abutments was measured using an Instron Universal load-testing machine.

RESULTS

The mean peak removal force for comparable abutments was significantly different ( p < 0.05): (1) where the screw access channel was completely filled with Memosil compared with those partially filled with Memosil; (2) with platform sizes--WP > RP > NP; (3) with alteration of axial wall height--1/3 removed > unadjusted = 2/3 removed > total wall removal.

CONCLUSIONS

The retention of castings cemented to implant abutments with TempBond is influenced by the wall height, platform size, and the filling modality of the screw access channel.

摘要

目的

影响固定修复体与天然基牙固位的因素已为人熟知。相比之下,关于影响固定修复体与种植体基台粘结固位的因素却知之甚少。本研究的目的是调查种植体基台壁高度、平台尺寸和螺钉通道填充方法的变化对使用TempBond粘结到种植体基台的铸件固位力的影响。

材料与方法

使用每种平台尺寸(窄型(NP)、常规型(RP)和宽型(WP))的四个15°预成角度基台(诺贝尔生物保健Replace Select Esthetic)。在每组基台中,螺钉通道轴向壁要么不做调整,要么去除三分之一、三分之二或全部去除。螺钉通道要么完全用乙烯基聚硅氧烷印模材料Memosil填充,要么部分填充。为每个基台制作一个带有可拆卸附件的铸件。铸件用TempBond粘结到基台上。使用Instron万能负载试验机测量将粘结的铸件与基台分离所需的拉力。

结果

可比基台的平均峰值去除力有显著差异(p < 0.05):(1)螺钉通道完全用Memosil填充的与部分用Memosil填充的相比;(2)平台尺寸——WP > RP > NP;(3)轴向壁高度改变时——去除三分之一 > 未调整 = 去除三分之二 > 完全去除壁。

结论

用TempBond粘结到种植体基台的铸件固位力受壁高度、平台尺寸和螺钉通道填充方式的影响。