教学诊所牙科设备水线系统中铜绿假单胞菌的存在情况。

The presence of Pseudomonas aeruginosa in the dental unit waterline systems of teaching clinics.

作者信息

Al-Hiyasat A S, Ma'ayeh S Y, Hindiyeh M Y, Khader Y S

机构信息

Department of Restorative Dentistry, Faculty of Dentistry, Jordan University of Science and Technology, Irbid, Jordan.

出版信息

Int J Dent Hyg. 2007 Feb;5(1):36-44. doi: 10.1111/j.1601-5037.2007.00221.x.

Abstract

OBJECTIVE

The objective of this study was to evaluate the extent of Pseudomonas aeruginosa contamination of Dental Unit Water (DUW) at a Dental Teaching Center in Jordan.

METHODS

Water samples were collected from 30 dental units, 10 from each of three teaching clinics, namely conservative dentistry, periodontology, and prosthodontics. Samples were collected from the outlet of the air/water syringe, high-speed handpiece and water cup filler, at the beginning of the working day (before use), after 2 min flushing, and at midday.

RESULTS

P. aeruginosa was detected in 86.7% (26/30) of the dental units at the beginning of the working day, and in 73.3% (22/30) after 2 min of flushing and at midday. Conservative dentistry units had the highest counts, followed by periodontology and prosthodontics (P<0.05). Overall, the highest counts (log10 count CFU ml-1) were at the beginning of the working day (1.38+/-1.05), and the lowest counts after flushing for 2 min (1.10+/-1.03), and higher numbers were seen again at midday (1.15+/-1.04) (P<0.05).

CONCLUSIONS

86.7% of the dental units were contaminated with P. aeruginosa, the conservative dentistry units had the highest amount of contamination. Flushing the DUW for 2 min significantly reduced the counts of P. aeruginosa.

摘要

目的

本研究的目的是评估约旦一家牙科教学中心牙科设备用水(DUW)中铜绿假单胞菌的污染程度。

方法

从30个牙科设备中采集水样,其中三个教学诊所各10个,分别为保守牙科、牙周病科和口腔修复科。在工作日开始时(使用前)、冲洗2分钟后和中午,从空气/水注射器出口、高速手机和水杯注水口采集样本。

结果

在工作日开始时,86.7%(26/30)的牙科设备检测到铜绿假单胞菌,冲洗2分钟后及中午时,这一比例为73.3%(22/30)。保守牙科设备的菌数最高,其次是牙周病科和口腔修复科(P<0.05)。总体而言,菌数最高(log10菌落形成单位CFU/ml)出现在工作日开始时(1.38±1.05),冲洗2分钟后最低(1.10±1.03),中午时又出现较高菌数(1.15±1.04)(P<0.05)。

结论

86.7%的牙科设备被铜绿假单胞菌污染,保守牙科设备的污染程度最高。将DUW冲洗2分钟可显著减少铜绿假单胞菌的菌数。