Gallagher L M, Owen L J, Keevil B G
Department of Clinical Biochemistry, South Manchester University Hospitals NHS Trust, Manchester M23 9LT, UK.
Ann Clin Biochem. 2007 Jan;44(Pt 1):48-56. doi: 10.1258/000456307779595922.
We aimed to develop a sensitive assay to quantitate serum concentrations of both androstenedione and testosterone within the female range simultaneously, using liquid chromatography-tandem mass spectrometry (LC-MS/MS) for use in the routine clinical laboratory and to compare this method with immunoassay.
Samples (200 microL) were prepared by liquid-liquid extraction using (1 mL) methyl-tert-butyl-ether. Deuterated androstenedione and testosterone were used as internal standards.
The standard curve was linear to 50 nmol/L, the lower limit of quantitation was 0.25 nmol/L, and intra- and inter-assay coefficients of variation were < 10% for both androgens over the range 0.3-35 nmol/L. There was a poor relationship between the LC-MS/MS and the radioimmunoassay methods for androstenedione with the LC-MS/MS generally giving lower results. For testosterone, the LC-MS/MS and immunoassay methods compared well at all concentrations. However, when female samples only were examined, the agreement deteriorated.
We have developed a sensitive and precise LC-MS/MS method, which gives more accurate results for all androstenedione measurements and low testosterone concentrations than immunoassay.
我们旨在开发一种灵敏的检测方法,使用液相色谱 - 串联质谱法(LC-MS/MS)同时定量女性范围内的血清雄烯二酮和睾酮浓度,以便用于常规临床实验室,并将该方法与免疫测定法进行比较。
样品(200微升)通过使用(1毫升)甲基叔丁基醚进行液 - 液萃取制备。氘代雄烯二酮和睾酮用作内标。
标准曲线在50纳摩尔/升范围内呈线性,定量下限为0.25纳摩尔/升,在0.3 - 35纳摩尔/升范围内,两种雄激素的批内和批间变异系数均<10%。LC-MS/MS与雄烯二酮放射免疫测定法之间的相关性较差,LC-MS/MS通常给出较低的结果。对于睾酮,LC-MS/MS和免疫测定法在所有浓度下比较良好。然而,仅检查女性样品时,一致性变差。
我们开发了一种灵敏且精确的LC-MS/MS方法,与免疫测定法相比,该方法在所有雄烯二酮测量以及低睾酮浓度测定中能给出更准确的结果。