Parekh Parag, Green W Richard, Stark Walter J, Akpek Esen Karamursel
Wilmer Eye Institute, Johns Hopkins University, Baltimore, Maryland 21287, USA.
Ophthalmology. 2007 Feb;114(2):232-7. doi: 10.1016/j.ophtha.2006.10.037.
To report the occurrence of subluxation of suture-fixated posterior chamber (PC) intraocular lenses (IOL) and elucidate the mechanisms involved.
Prospective clinicopathologic study.
A single 10-0 Prolene suture explanted from a patient who experienced subluxation of his PC-IOL, 11.5 years after placement. Furthermore, multiple 10-0 Prolene sutures and PC-IOLs used for iris fixation were studied as controls.
Scanning electron microscopy (SEM) was used to analyze the surface of the explanted suture. In addition, randomly selected 10-0 Prolene sutures cut with Vannas scissors and cut with the positioning holes of a randomly selected PC-IOL identical to that implanted in the patient's eye were examined as controls. Finally, the positioning holes of several randomly selected, iris-fixated PC-IOLs were studied using SEM with particular attention to surface quality and edge finish.
Presence of any signs of suture degradation, the character of the cut edge of the suture, as well as the characteristics of the positioning holes of the PC-IOLs.
Scanning electron microscopy of the explanted suture revealed sharply cut edges, without significant degradation of the suture, and no intact loop. Scanning electron microscopy of the control suture cut with a PC-IOL demonstrated a similarly cut edge. The positioning holes of the examined PC-IOLs had a sharp edge, and some also had an imperfect finish.
We conclude that the surface properties of the positioning holes lead to cutting of the suture, and subsequent subluxation of the PC-IOL.
报告缝线固定后房型人工晶状体(IOL)半脱位的发生情况,并阐明其相关机制。
前瞻性临床病理研究。
从一名人工晶状体植入11.5年后发生半脱位的患者眼中取出的一根10-0聚丙烯缝线。此外,还研究了多根用于虹膜固定的10-0聚丙烯缝线和后房型人工晶状体作为对照。
使用扫描电子显微镜(SEM)分析取出缝线的表面。此外,将随机选择的用Vannas剪刀剪断的10-0聚丙烯缝线,以及用与植入患者眼中相同的随机选择的后房型人工晶状体的定位孔剪断的缝线作为对照进行检查。最后,使用扫描电子显微镜研究几根随机选择的虹膜固定后房型人工晶状体的定位孔,特别关注表面质量和边缘光洁度。
缝线降解的任何迹象、缝线切割边缘的特征以及后房型人工晶状体定位孔的特征。
对取出的缝线进行扫描电子显微镜检查发现边缘切割整齐,缝线无明显降解,且无完整的环。用后房型人工晶状体剪断的对照缝线的扫描电子显微镜检查显示切割边缘类似。所检查的后房型人工晶状体的定位孔边缘锋利,有些边缘光洁度也欠佳。
我们得出结论,定位孔的表面特性导致缝线切割,进而导致后房型人工晶状体半脱位。