Furui S, Yamauchi T, Makita K, Takeshita K, Irie T, Tsuchiya K, Sawada S, Nakamura H, Okazaki M
Department of Radiology, National Defense Medical College, Saitama, Japan.
Radiology. 1992 Jan;182(1):283-4. doi: 10.1148/radiology.182.1.1727299.
A foreign body retrieval system composed of a three-lumen catheter, a guide wire, and a wire loop nearly perpendicular to the axis of the catheter was used to retrieve catheter fragments located in the inferior vena cava in two patients and in the superior vena cava in one patient. Fragments were looped within 1 minute of catheterization and were retrieved without complication. Catheter fragments placed in the pulmonary artery of one dog were also successfully retrieved.
一种由三腔导管、导丝和一个几乎垂直于导管轴线的金属丝环组成的异物取出系统,用于取出两名患者下腔静脉和一名患者上腔静脉中的导管碎片。碎片在插管后1分钟内被套住,并成功取出,无并发症。置于一只狗肺动脉中的导管碎片也被成功取出。