使用组织等效正比计数器进行校准和本底测量。

Calibration and background measurements with a tissue equivalent proportional counter.

作者信息

Autischer M, Beck P, Kindl P, Latocha M, Rollet S

机构信息

ARC Seibersdorf Research GmbH, 2444 Seibersdorf, Austria.

出版信息

Radiat Prot Dosimetry. 2007;125(1-4):429-32. doi: 10.1093/rpd/ncl557. Epub 2007 Feb 3.

Abstract

A tissue-equivalent-proportional counter (TEPC) instrument has been used as the reference instrument for cosmic radiation measurement at flight altitudes by several institutes. For purposes of characterisation the response of the instrument has been investigated under different standard radiation conditions, in terms of radiation particle, energy and angle of incidence. Photon sources and photon beams of energies up to 6.6 MeV and neutron beams up to 200 MeV were used. To have a better understanding of the shielding influence of the instrument assembly, the angle dependence of response was analysed for several radiation conditions. Specific measurement conditions were simulated with the Monte Carlo transport code, FLUKA. The measured instrument response was compared with simulation results. It was demonstrated, that simulations were very helpful to understand the instrument's response. The TEPC instrument used by the Austrian Research Centre Seibersdorf (ARCS) research simulates the energy deposition in a unit density tissue volume of 2 microm diameter, of similar size to a cell nucleus. Pure propane at low pressure is used as measurement gas. To characterise the instrument at low dose rates background measurements were done 800 m below ground and at the ultra low level laboratory in Gran Sasso, 1380 m below ground. These results were compared with measurements on the Earth's surface at different altitudes on mountains up to 3480 m above sea level. The significant increase of the expected dose rate is well reproduced by the experiments at mountain altitudes. As a result of this study a full characterisation and a thorough understanding of the performance and reliability of the detector was achieved.

摘要

几个机构已将组织等效正比计数器(TEPC)仪器用作飞行高度宇宙辐射测量的参考仪器。为了进行特性描述,已在不同的标准辐射条件下,根据辐射粒子、能量和入射角对该仪器的响应进行了研究。使用了能量高达6.6 MeV的光子源和光子束以及能量高达200 MeV的中子束。为了更好地理解仪器组件的屏蔽影响,分析了几种辐射条件下响应的角度依赖性。使用蒙特卡罗传输代码FLUKA模拟了特定的测量条件。将测量的仪器响应与模拟结果进行了比较。结果表明,模拟对于理解仪器的响应非常有帮助。奥地利塞伯斯多夫研究中心(ARCS)使用的TEPC仪器模拟了直径为2微米的单位密度组织体积中的能量沉积,其大小与细胞核相似。低压纯丙烷用作测量气体。为了在低剂量率下对仪器进行特性描述,在地下800米处以及在格兰萨索地下1380米处的超低本底实验室进行了本底测量。将这些结果与在海拔高达3480米的不同山区地面上的测量结果进行了比较。山区实验很好地再现了预期剂量率的显著增加。通过这项研究,实现了对探测器性能和可靠性的全面特性描述和深入理解。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验