Esophageal electrodes allow precise assessment of cardiac output by bioimpedance.

作者信息

Balestra B, Malacrida R, Leonardi L, Suter P, Marone C

机构信息

Centro Cure Intense, Ospedale San Giovanni, Bellinzona, Switzerland.

出版信息

Crit Care Med. 1992 Jan;20(1):62-7. doi: 10.1097/00003246-199201000-00018.

Abstract

OBJECTIVES

To analyze the impact of the position of the thoracic external electrodes on the values of cardiac output measured by electrical bioimpedance and to compare the results obtained by bioimpedance with those values determined by thermodilution in critically ill patients.

DESIGN

Open, prospective, comparative trial.

SETTING

ICU of a teaching hospital.

PATIENTS

Twenty healthy volunteers and ten critically ill patients.

INTERVENTIONS

Measurements of cardiac output by bioimpedance at rest and after physical activity in normal volunteers and after changing the neck or xiphoid electrodes. Comparisons of cardiac output obtained by thermodilution and bioimpedance with internal and external electrodes in patients.

MEASUREMENTS AND MAIN RESULTS

Mean +/- SD values are presented. Cardiac output values at rest and after exercise were 6.7 +/- 1.3 and 10.8 +/- 2.6 L/min at rest and after exercise, respectively (p less than .001). Displacement of the xiphoid electrodes 3 cm in the caudal direction was accompanied by a decrease of the mean cardiac output from 7.1 +/- 1.2 to 5.8 +/- 1.3 L/min (p less than .001) and displacement 3 and 6 cm cranially was accompanied by increases in cardiac output from 7.1 +/- 1.2 to 8.1 +/- 1.4 L/min (p less than .001) and 8.6 +/- 1.5 L/min (p less than .001), respectively. In the ten patients, cardiac output measurements were virtually identical when results obtained by thermodilution (6.7 +/- 3.1 L/min) were compared with those results obtained by bioimpedance using internal esophageal (6.6 +/- 3.1 L/min), but not external (4.7 +/- 1.6 L/min) electrodes.

CONCLUSIONS

a) The values of cardiac output derived from measurements obtained by bioimpedance using internal electrodes were comparable with those values derived from thermodilution. b) Values of cardiac output from bioimpedance studies with external electrodes were dependent on the position of the xiphoid electrodes.