Andreotti Felicita, Lavorgna Alberto, Coluzzi Giulio, Vincenzoni Claudio, Rebuzzi Antonio G, Porto Italo, Crea Filippo
Int J Cardiol. 2007 Apr 12;117(1):e1-3. doi: 10.1016/j.ijcard.2006.07.027. Epub 2007 Feb 9.
We report an unusual case of breakage and embolisation of a femoral access closure device (Angio-Seal), which caused acute leg ischemia after percutaneous coronary intervention in a diabetic patient with documented peripheral vascular disease. Embolectomy from the superficial femoral artery revealed the polymer anchor split in two parts and embedded within a large (25x5 mm) thrombus.