Zhang Guodong, Terry Alvin V, Bartlett Michael G
Department of Pharmaceutical and Biomedical Sciences, College of Pharmacy, The University of Georgia, Athens, GA 30602-2352, USA.
Rapid Commun Mass Spectrom. 2007;21(6):920-8. doi: 10.1002/rcm.2914.
A simple, sensitive and rapid liquid chromatography/electrospray ionization tandem mass spectrometry (LC/ESI-MS/MS) method was developed and validated for simultaneous quantification of olanzapine, clozapine, ziprasidone, haloperidol, risperidone, and its active metabolite 9-hydroxyrisperidone, in rat plasma using midazolam as internal standard (IS). The analytes were extracted from rat plasma using a single step liquid-liquid extraction technique. The compounds were separated on a Waters Atlantis dC-18 (30 mm x 2.1 mm i.d., 3 microm) column using a mobile phase of acetonitrile/5 mM ammonium formate (pH 6.1 adjusted with formic acid) with gradient elution. All of the analytes were detected in positive ion mode using multiple reaction monitoring (MRM). The method was validated and the specificity, linearity, lower limit of quantitation (LLOQ), precision, accuracy, recoveries and stability were determined. LLOQ was 0.1 ng/mL and correlation coefficient (R(2)) values for the linear range of 0.1-100 ng/mL were 0.997 or greater for all the analytes. The intra-day and inter-day precision and accuracy were better than 8.05%. The relative and absolute recovery was above 77% and matrix effects were low for all the analytes except for ziprasidone. This validated method has been successfully used to quantify the plasma concentration of the analytes after chronic treatment with antipsychotic drugs.
建立了一种简单、灵敏且快速的液相色谱/电喷雾电离串联质谱法(LC/ESI-MS/MS),并进行了验证,该方法以咪达唑仑作为内标(IS),可同时定量大鼠血浆中的奥氮平、氯氮平、齐拉西酮、氟哌啶醇、利培酮及其活性代谢物9-羟基利培酮。采用单步液液萃取技术从大鼠血浆中提取分析物。使用乙腈/5 mM甲酸铵(用甲酸调节pH至6.1)作为流动相并进行梯度洗脱,在Waters Atlantis dC-18(30 mm×2.1 mm内径,3 µm)柱上分离化合物。所有分析物均采用多反应监测(MRM)在正离子模式下进行检测。对该方法进行了验证,并测定了特异性、线性、定量下限(LLOQ)、精密度、准确度、回收率和稳定性。LLOQ为0.1 ng/mL,所有分析物在0.1 - 100 ng/mL线性范围内的相关系数(R²)值均为0.997或更高。日内和日间精密度及准确度均优于8.05%。除齐拉西酮外,所有分析物的相对回收率和绝对回收率均高于77%,基质效应较低。该经过验证的方法已成功用于定量抗精神病药物长期治疗后分析物的血浆浓度。