Horáková Jana, Petr Jan, Maier Vítezslav, Znaleziona Joanna, Stanová Andrea, Marák Jozef, Kaniansky Dusan, Sevcík Juraj
Department of Analytical Chemistry, Palacký University, Trída Svobody 8, Olomouc, Czech Republic.
J Chromatogr A. 2007 Jul 6;1155(2):193-8. doi: 10.1016/j.chroma.2007.01.075. Epub 2007 Jan 28.
An on-line preconcentration capillary electrophoresis (CE) technique, which combines a large volume sample stacking with a dynamic pH junction technique, is introduced in this paper. This dynamic pH junction with co-electroosmotic migration is formed between sodium borate pH 9.5 and sodium phosphate pH 2.5 with 150 mM sodium dodecylsulfate (SDS). A full capillary based injection allows determination of weak acidic compounds at ppb concentration levels (achieved LOD for benzoic acid was 11 nmol L(-1)). The proposed preconcentration method was compared with ITP/ITP (LOD 120 nmol L(-1)), ITP/CZE (LOD 740 nmol L(-1)) and a simple CZE method (LOD 23,330 nmol L(-1)). The analytical potential of this method was assessed with juice test samples.