Tseng Yen-Jia, Tsai Yu-Duan, Jiang Shiuh-Jen
Department of Chemistry, National Sun Yat-sen University, Kaohsiung 80424, Taiwan.
Anal Bioanal Chem. 2007 Apr;387(8):2849-55. doi: 10.1007/s00216-007-1143-0. Epub 2007 Feb 13.
Slurry sampling electrothermal vaporization dynamic reaction cell inductively coupled plasma mass spectrometry (ETV-DRC-ICP-MS) has been applied to determine Fe, Co, Ni, Cu, and Zn in biological samples. The influences of instrument operating conditions and slurry preparation on the ion signals were reported. Pd was used as the modifier. The effectiveness of the ETV sample introduction technique and dynamic reaction cell in alleviating various spectral interferences in ICP-MS analysis has been demonstrated. This method has been applied to determine Fe, Co, Ni, Cu, and Zn in NIST SRM 1573a tomato leaves reference material and NRCC DORM-2 dogfish muscle reference material and also real samples such as a tea and a swordfish sample purchased locally. Since the sensitivities of the elements studied in slurry and aqueous solution were different, an analyte addition technique was used for the determinations. The analytical results of the reference materials agreed with the certified values. The precision between sample replicates was better than 6% for all determinations. The method detection limit estimated from analyte addition curves was 0.01, 0.006, 0.007, 0.004, and 0.006 microg g(-1) for Fe, Co, Ni, Cu, and Zn, respectively, in the original biological samples.
浆液进样电热蒸发动态反应池电感耦合等离子体质谱法(ETV-DRC-ICP-MS)已用于测定生物样品中的铁、钴、镍、铜和锌。报道了仪器操作条件和浆液制备对离子信号的影响。钯用作改性剂。已证明ETV进样技术和动态反应池在减轻ICP-MS分析中各种光谱干扰方面的有效性。该方法已用于测定NIST SRM 1573a番茄叶标准物质和NRCC DORM-2鲨鱼肌肉标准物质以及本地购买的茶叶和箭鱼等实际样品中的铁、钴、镍、铜和锌。由于所研究元素在浆液和水溶液中的灵敏度不同,测定时采用了分析物添加技术。标准物质的分析结果与认定值相符。所有测定中,样品重复之间的精密度优于6%。根据分析物添加曲线估算,原始生物样品中铁、钴、镍、铜和锌的方法检出限分别为0.01、0.006、0.007、0.004和0.006 μg g⁻¹。