Spotnitz H M, Ott G Y, Bigger J T, Steinberg J S, Livelli F
Department of Surgery, Columbia-Presbyterian Medical Center, New York, New York 10032.
Ann Thorac Surg. 1992 Feb;53(2):253-7. doi: 10.1016/0003-4975(92)91327-6.
Experience with combined transvenous pacemaker and implantable cardioverter-defibrillator insertion in 21 patients is described. Special techniques are needed to avoid potentially lethal pacemaker-implantable cardioverter-defibrillator interaction. Separation between leads for the two devices should be maximized. The electrophysiologic criteria for successful device function can be met, even when some leads for both devices must be placed by a transvenous approach.
本文描述了21例患者联合植入经静脉起搏器和植入式心脏复律除颤器的经验。需要特殊技术来避免潜在致命的起搏器-植入式心脏复律除颤器相互作用。应尽量增大两种设备导线之间的间距。即使两种设备的一些导线都必须通过经静脉途径放置,也可以满足设备正常运行的电生理标准。