Suppr超能文献

Rigid-flexible outer sheath model using slider linkage locking mechanism and air pressure for endoscopic surgery.

作者信息

Yagi Akihiko, Matsumiya Kiyoshi, Masamune Ken, Liao Hongen, Dohi Takeyoshi

机构信息

Graduate School of Information Science and Techonology, The University of Tokyo, 7-3-1 Hongo Bunkyo-ku, Tokyo, 133-8656, Japan.

出版信息

Med Image Comput Comput Assist Interv. 2006;9(Pt 1):503-10. doi: 10.1007/11866565_62.

Abstract

The objective of this paper is to develop an outer sheath for flexible endoscopic manipulators. This sheath can switch two states including flexible and rigid, and make a rigid curved path for inserting manipulators. The flexible mode can be curved into a required shape. The rigid mode can hold the shape of the sheath, and then keep the path for instruments. Through the managed path, the flexible manipulators become easy to reach the target. We proposed a serial multi joint model to realize the flexible mechanism. This model is composed of a set of frame units which are connected serially. Each unit can be rotated to a given angle around the center of the joint. We developed a slider-link mechanism and a gear stopper controlled by air pressure for rigid mode. We designed and fabricated the prototype with a diameter of 16 mm and length of 290 mm. The experiment showed that the device could be switched from the flexible mode to the rigid mode when the air pressure was over 150 kPa, and each joint could hold its angle against the maximum 400 mNm. The phantom experiment showed that the flexible devices are possible to transmit the wire tension to the endpoint of the manipulator without changing the curving shape with by the developed outer sheath device.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验