Krejcová Anna, Cernohorský Tomás, Pouzar Miloslav
Department of Environmental Protection, University of Pardubice, Pardubice, Czech Republic.
Anal Chim Acta. 2007 Jan 23;582(2):208-13. doi: 10.1016/j.aca.2006.09.031. Epub 2006 Sep 26.
A simple analytical method sufficient for the determination of Na, K, Ca, Mg, Al, B, Ba, Cd, Co, Cr, Cu, Fe, Li, Mn, Mo, Ni, Pb, Sr, V and Zn at mg kg(-1) levels in pure NaOH, KOH, NaCl and KCl using the optical emission spectrometry was developed. The results of direct determination with multi-elemental aqueous standards were compared with them obtained by the internal standardisation, by the standard addition methods and with the maximal allowable contents of above-mentioned elements in pure chemicals. The method was shown to be very sensitive and exhibits following limits of detection: Na 0.90, K 3.0, Ca 0.064, Mg 0.026, Al 0.43, B 0.13, Ba 0.015, Cd 0.023, Co 0.056, Cr 0.041, Cu 0.063, Fe 0.060, Li 0.017, Mn 0.035, Mo 0.19, Ni 0.055, Pb 0.39, Sr 0.030, V 0.065 and Zn 0.043 (all in mg kg(-1)). The method presents a satisfactory precision (relative standard deviation 4-11%), high analytical recoveries, linear responses of at least four orders of magnitude, accuracy and low contamination susceptibility.
建立了一种简单的分析方法,该方法足以使用光发射光谱法测定纯氢氧化钠、氢氧化钾、氯化钠和氯化钾中毫克每千克(mg kg(-1))水平的钠、钾、钙、镁、铝、硼、钡、镉、钴、铬、铜、铁、锂、锰、钼、镍、铅、锶、钒和锌。将多元素水溶液标准品直接测定的结果与通过内标法、标准加入法获得的结果以及纯化学品中上述元素的最大允许含量进行了比较。结果表明该方法非常灵敏,检测限如下:钠0.90、钾3.0、钙0.064、镁0.026、铝0.43、硼0.13、钡0.015、镉0.023、钴0.056、铬0.041、铜0.063、铁0.060、锂0.017、锰0.035、钼0.19、镍0.055、铅0.39、锶0.030、钒0.065和锌0.043(均以mg kg(-1)计)。该方法具有令人满意的精密度(相对标准偏差为4-11%)、高分析回收率、至少四个数量级的线性响应、准确性和低污染敏感性。