对接受碘I 131治疗的猫的辐射剂量率测量中两种仪器之间一致性的评估。

Evaluation of agreement between two instruments in measurements of radiation dose rates in cats that underwent iodine I 131 treatment.

作者信息

Chalmers Heather J, Scrivani Peter V, Dykes Nathan L, Hubble Lorna J, Hobbs Jeff M, Erb Hollis N

机构信息

Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY 14853, USA.

出版信息

Am J Vet Res. 2007 Apr;68(4):354-7. doi: 10.2460/ajvr.68.4.354.

Abstract

OBJECTIVE

To assess agreement between a commercially available Geiger-Meuller (GM) survey meter and millirem tissue-equivalent (TE) meter for measuring radioactivity in cats treated with sodium iodine I 131 ((131)I).

ANIMALS

15 cats with hyperthyroidism and undergoing (131)I treatment.

PROCEDURES

Duplicate measurements were obtained at a distance of 30 cm from the thyroid region of each cat's neck by 2 observers who used both meters on day 3 or 5 after(131)I administration. Comparisons of measurements between meters and observers were made, with limits of agreement defined as the mean difference +/- 2 SDs of the differences.

RESULTS

For observer 1, the mean of the differences in the 2 meters' measurements in all cats was 0.012 mSv/h (SD, 0.011 mSv/h). For observer 2, the mean of the differences in measurements was 0.012 mSv/h (SD, 0.010 mSv/h). For the GM meter, the mean of the differences of the 2 observers for all cats was 0.003 mSv/h (SD, 0.011 mSv/h). For the TE meter, the mean of the differences of the 2 observers for all cats was 0.003 mSv/h (SD, 0.007 mSv/h).

CONCLUSIONS AND CLINICAL RELEVANCE

Results indicate that there was considerable agreement between meters and observers in measurements of radioactivity in cats treated with (131)I. Measurements obtained by use of the GM meter may be approximately 0.01 mSv/h less than or 0.03 mSv/h higher than those obtained with the TE meter. If this range is acceptable for an institution's release criteria, the 2 meters should be considered interchangeable and acceptable for clinical use.

摘要

目的

评估一种商用盖革 - 米勒(GM)测量仪与毫雷姆组织等效(TE)测量仪在测量接受碘I 131(¹³¹I)治疗的猫的放射性方面的一致性。

动物

15只患有甲状腺功能亢进且正在接受¹³¹I治疗的猫。

步骤

两名观察者在¹³¹I给药后第3天或第5天,使用两种测量仪在距每只猫颈部甲状腺区域30厘米处进行重复测量。对测量仪之间以及观察者之间的测量结果进行比较,一致性界限定义为差异的均值±2个标准差。

结果

对于观察者1,所有猫中两种测量仪测量结果的差异均值为0.012毫希沃特/小时(标准差,0.011毫希沃特/小时)。对于观察者2,测量结果的差异均值为0.012毫希沃特/小时(标准差,0.010毫希沃特/小时)。对于GM测量仪,所有猫中两名观察者测量结果的差异均值为0.003毫希沃特/小时(标准差,0.011毫希沃特/小时)。对于TE测量仪,所有猫中两名观察者测量结果的差异均值为0.003毫希沃特/小时(标准差,0.007毫希沃特/小时)。

结论及临床意义

结果表明,在测量接受¹³¹I治疗的猫的放射性方面,测量仪与观察者之间存在相当程度的一致性。使用GM测量仪获得的测量结果可能比使用TE测量仪获得的结果低约0.01毫希沃特/小时或高0.03毫希沃特/小时。如果该范围符合机构的放行标准,则应认为这两种测量仪可互换且适用于临床使用。