Quantitative computed tomography: comparative study using different scanners with two calibration phantoms.

作者信息

Suzuki S, Yamamuro T, Okumura H, Yamamoto I

机构信息

Department of Orthopaedic Surgery, Faculty of Medicine, Kyoto University, Japan.

出版信息

Br J Radiol. 1991 Nov;64(767):1001-6. doi: 10.1259/0007-1285-64-767-1001.

Abstract

To assess computed tomography (CT) scanners in vertebral quantitative computed tomography (QCT) measurement, five cadaveric vertebrae fixed in a water phantom were measured using 16 CT scanners of 10 different models using two types (CaCO3 and K2HPO4) of reference phantom. Although the same reference phantoms were used, the QCT values varied markedly depending on the CT scanner employed. The differences in QCT values were greater in the equivalent amount of CaCO3 than in that of K2HPO4. The largest difference between CT scanners was 40 mg/cm3 in the equivalent amount of CaCO3 and 28 mg/cm3 in that of K2HPO4. These differences were reflected among CT scanners by different slopes and intercepts of the reference lines of CaCO3, K2HPO4 and ash density depending on the CT scanner used. However, good correlations (r greater than 0.97) were observed for the QCT values obtained by various CT scanners. These results show that QCT values on one machine can be compared more closely with those of another machine if conversion formulae, determined by using materials equivalent to human vertebral bone, are employed.

摘要

为评估计算机断层扫描(CT)扫描仪在椎体定量计算机断层扫描(QCT)测量中的性能,使用10种不同型号的16台CT扫描仪,对固定在水模体中的5个尸体椎体进行测量,使用两种类型(碳酸钙和磷酸氢二钾)的参考模体。尽管使用了相同的参考模体,但QCT值根据所使用的CT扫描仪的不同而有显著差异。碳酸钙等量时QCT值的差异大于磷酸氢二钾等量时的差异。CT扫描仪之间的最大差异在碳酸钙等量时为40mg/cm³,在磷酸氢二钾等量时为28mg/cm³。这些差异在不同CT扫描仪中表现为,根据所使用的CT扫描仪不同,碳酸钙、磷酸氢二钾和灰密度参考线的斜率和截距也不同。然而,不同CT扫描仪获得的QCT值之间观察到良好的相关性(r大于0.97)。这些结果表明,如果采用由等同于人体椎骨的材料确定的转换公式,一台机器上的QCT值可以与另一台机器上的QCT值进行更紧密的比较。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验