Accuracy, Linearity and Precision of Spine QCT vBMD Phantom Measurements for Different Brands of CT Scanner: A Multicentre Study.

作者信息

Zhao Yingwei, Li Kai, Duanmu Yangyang, Wang Ling, Xu Xiaoming, Zhang Yong, Tang Jing, Zhang Yujing, Li Zhenlin, Hind Karen, Blake Glen M, Cheng Xiaoguang

机构信息

Department of Radiology, Beijing Jishuitan Hospital, Beijing, China.

Department of Radiology, First Affiliated Hospital, University of Science and Technology of China, Hefei, China.

出版信息

J Clin Densitom. 2022 Jan-Mar;25(1):34-42. doi: 10.1016/j.jocd.2021.02.004. Epub 2021 Feb 28.

Abstract

We describe a multicenter study using the European Spine Phantom (ESP) to compare the accuracy, linearity and precision of QCT measurements of spine vBMD between different brands of scanner, different models of the same brand and identical units of the same model. Ten scans of the same ESP with repositioning were performed on forty CT scanners from five manufacturers in different hospitals across China, all calibrated with the Mindways QCT system. The three ESP vertebral bodies simulating low (L1), medium (L2) and high (L3) vBMD and their average (L1-3 vBMD) were compared with phantom values. Linearity was assessed using the standard error of the estimate derived from linear regression. Precision errors were expressed as the standard deviation of the ten measurements on each scanner. Median (IQR) vBMD over all forty CT scanners compared with phantom values were: L1: 52.2 (49.9-56.4) vs 51.0; L2: 104.4 (101.2-108.6) vs 102.2; L3: 201.4 (195.0-204.9) vs 200.4; L1-3: 119.3 (116.6-123.2) vs 117.9 mg/cm. Statistically significant differences in L1-3 vBMD were found between different brands (p= 0.005) and between different models of the same brand and identical units of the same model (both p< 0.001). Cross-calibration using linear regression gave a good fit for all forty systems with a median standard error of the estimate of 1.7 mg/cm. The median precision error for L1-3 vBMD was 0.61 mg/cm. Statistically significant differences in spine vBMD measurements between different scanners reinforce the importance of cross-calibration in multi-center studies. Cross-calibration can be reliably performed using linear regression equations.

摘要

我们描述了一项多中心研究,该研究使用欧洲脊柱体模(ESP)来比较不同品牌扫描仪、同一品牌不同型号以及同一型号相同设备之间脊柱骨密度(vBMD)的定量CT(QCT)测量的准确性、线性度和精密度。在中国不同医院的五家制造商的40台CT扫描仪上,对同一ESP进行了十次重新定位扫描,所有扫描仪均使用Mindways QCT系统进行校准。将模拟低(L1)、中(L2)和高(L3)vBMD的三个ESP椎体及其平均值(L1-3 vBMD)与体模值进行比较。使用线性回归得出的估计标准误差评估线性度。精密度误差表示为每台扫描仪上十次测量的标准差。与体模值相比,所有40台CT扫描仪上的中位数(四分位间距)vBMD为:L1:52.2(49.9-56.4)对51.0;L2:104.4(101.2-108.6)对102.2;L3:201.4(195.0-204.9)对200.4;L1-3:119.3(116.6-123.2)对117.9mg/cm³。在不同品牌之间(p = 0.005)以及同一品牌不同型号和同一型号相同设备之间(均p < 0.001)发现L1-3 vBMD存在统计学显著差异。使用线性回归进行交叉校准对所有40个系统都有很好的拟合,估计的中位数标准误差为1.7mg/cm³。L1-3 vBMD的中位数精密度误差为0.61mg/cm³。不同扫描仪之间脊柱vBMD测量的统计学显著差异强化了多中心研究中交叉校准的重要性。使用线性回归方程可以可靠地进行交叉校准。