Critical noise level reduction for correct identification of late potentials: an illustrative case report.

作者信息

Kulakowski P, Murgatroyd F D, Camm A J

机构信息

Department of Cardiological Sciences, St. George's Hospital Medical School, London, England.

出版信息

Clin Cardiol. 1991 Sep;14(9):779-83. doi: 10.1002/clc.4960140915.

Abstract

The precision of detection of late potentials (LPs) on signal-averaged electrocardiography (SAECG) depends on the degree of noise reduction. The recommended low-noise endpoint is 0.3 microV. We describe a patient with sustained ventricular tachycardia after myocardial infarction with very low amplitude LPs in whom a noise level lower than 0.3 microV was necessary in order to obtain an abnormal result of the SAECG. The presence of LPs was confirmed by spectral temporal mapping of the SAECG.