Richert John D, Hogstrom Kenneth R, Fields Robert S, Matthews Kenneth L, Boyd Robert A
Department of Physics and Astronomy, Louisiana State University, 202 Nicholson Hall, Baton Rouge, LA 70803-4001, USA.
Phys Med Biol. 2007 May 7;52(9):2459-81. doi: 10.1088/0031-9155/52/9/009. Epub 2007 Apr 10.
The purpose of the present study is to demonstrate that the use of an electron applicator with energy-dependent source-to-collimator distances (SCDs) will significantly improve the dose homogeneity for abutted electron fields in segmented-field electron conformal therapy (ECT). Multiple Coulomb scattering theory was used to calculate and study the P(80-20) penumbra width of off-axis dose profiles as a function of air gap and depth. Collimating insert locations with air gaps (collimator-to-isocenter distance) of 5.0, 7.5, 11.5, 17.5 and 19.5 cm were selected to provide equal P(80-20) at a depth of 1.5 cm in water for energies of 6, 9, 12, 16 and 20 MeV, respectively, for a Varian 2100EX radiation therapy accelerator. A 15 x 15 cm(2) applicator was modified accordingly, and collimating inserts used in the variable-SCD applicator for segmented-field ECT were constructed with diverging edges using a computer-controlled hot-wire cutter, which resulted in 0.27 mm accuracy in the abutted edges. The resulting electron beams were commissioned for the pencil-beam algorithm (PBA) on the Pinnacle(3) treatment planning system. Four hypothetical planning target volumes (PTVs) and one patient were planned for segmented-field ECT using the new variable-SCD applicator, and the resulting dose distributions were compared with those calculated for the identical plans using the conventional 95 cm SCD applicator. Also, a method for quality assurance of segmented-field ECT dose plans using the variable-SCD applicator was evaluated by irradiating a polystyrene phantom using the treatment plans for the hypothetical PTVs. Treatment plans for all four of the hypothetical PTVs using the variable-SCD applicator showed significantly improved dose homogeneity in the abutment regions of the segmented-field ECT plans. This resulted in the dose spread (maximum dose-minimum dose), sigma, and D(90-10) in the PTV being reduced by an average of 32%, 29% and 32%, respectively. Reductions were most significant for abutted fields of nonadjacent energies. Planning segmented-field ECT using the variable-SCD applicator for a patient with recurrent squamous cell carcinoma of the left ear showed the dose spread, sigma, and D(90-10) of the dose distribution in the PTV being reduced by an average of 38%, 22% and 22%, respectively. The measured and calculated dose in a polystyrene phantom resulting from the variable-SCD, segmented-field ECT plans for the hypothetical PTVs showed good agreement; however, isolated differences between dose calculation and measurement indicated the need for a more accurate dose algorithm than the PBA for segmented-field ECT. These results confirmed our hypothesis that using the variable-SCD applicator for segmented-field ECT results in the PTV dose distribution becoming more homogenous and being within the range of 85-105% of the 'given dose'. Clinical implementation of this method requires variable-SCD applicators, and the design used in the present work should be acceptable, as should our methods for construction of the inserts. Dose verification measurements in a polystyrene phantom and the recommended improvements in dose calculation should be appropriate for quality assurance of segmented-field ECT.
本研究的目的是证明在分段野电子适形放疗(ECT)中,使用源到准直器距离(SCD)与能量相关的电子施源器将显著改善相邻电子野的剂量均匀性。采用多次库仑散射理论来计算和研究离轴剂量分布的P(80 - 20)半值层宽度与气隙和深度的函数关系。对于瓦里安2100EX放射治疗加速器,分别选择气隙(准直器到等中心距离)为5.0、7.5、11.5、17.5和19.5 cm的准直插入件位置,以便在水中1.5 cm深度处为6、9、12、16和20 MeV的能量提供相等的P(80 - 20)。相应地对一个15×15 cm²的施源器进行了改进,并使用计算机控制的热线切割机制作了用于分段野ECT的可变SCD施源器中的准直插入件,其对接边缘的精度为0.27 mm。在Pinnacle(3)治疗计划系统上对所得电子束进行铅笔束算法(PBA)调试。使用新型可变SCD施源器对四个假设的计划靶区(PTV)和一名患者进行分段野ECT计划,并将所得剂量分布与使用传统95 cm SCD施源器对相同计划计算的剂量分布进行比较。此外,通过使用假设PTV的治疗计划对聚苯乙烯模体进行照射,评估了使用可变SCD施源器进行分段野ECT剂量计划的质量保证方法。使用可变SCD施源器对所有四个假设PTV的治疗计划显示,在分段野ECT计划的对接区域剂量均匀性有显著改善。这导致PTV中的剂量展宽(最大剂量 - 最小剂量)、标准差和D(90 - 10)分别平均降低了32%、29%和32%。对于非相邻能量的相邻野,降低最为显著。对一名左耳复发性鳞状细胞癌患者使用可变SCD施源器进行分段野ECT计划,结果显示PTV中剂量分布的剂量展宽、标准差和D(90 - 10)分别平均降低了38%、22%和22%。可变SCD分段野ECT计划对假设PTV在聚苯乙烯模体中测量和计算的剂量显示出良好的一致性;然而,剂量计算与测量之间的孤立差异表明,对于分段野ECT,需要一种比PBA更精确的剂量算法。这些结果证实了我们的假设,即对于分段野ECT,使用可变SCD施源器会使PTV剂量分布变得更加均匀,且在“给定剂量”的85 - 105%范围内。该方法的临床应用需要可变SCD施源器,本研究中使用的设计以及我们制作插入件的方法应该是可接受的。在聚苯乙烯模体中的剂量验证测量以及推荐的剂量计算改进对于分段野ECT的质量保证应该是合适的。