术中使用Otologics全植入式听力装置进行听骨加载。

Intraoperative ossicular loading with the Otologics fully implantable hearing device.

作者信息

Jenkins Herman A, Pergola Nicholas, Kasic James

机构信息

Department of Otolaryngology, University of Colorado, Denver, CO, USA.

出版信息

Acta Otolaryngol. 2007 Apr;127(4):360-4. doi: 10.1080/00016480601089424.

Abstract

CONCLUSIONS

Use of the recently introduced intraoperative loading instrumentation during surgery improves the coupling efficiency and consistency of a fully implantable hearing device to the ossicles and ultimately leads to better patient performance with the device.

BACKGROUND

Patient performance with an implantable hearing device is dependent on effective coupling of the implant to the middle ear ossicles. New intraoperative instrumentation aids the surgeon in coupling the implantable hearing device and consistently optimizes patient performance.

MATERIALS AND METHODS

The middle ear transfer function for the Otologics MET Ossicular Stimulator (METF(MET)) is a measure of transducer coupling efficiency, defined as acoustic threshold (in dB HL) minus implant threshold (in dB MET). This transfer function was measured in the Otologics US Phase I Semi-Implantable Clinical Trial without intraoperative loading instrumentation and in the Otologics US Phase I Fully Implantable Clinical Trial with intraoperative loading instrumentation. The same operative techniques and transducer design were used in both studies. The theory of operation and surgical technique for using this instrumentation are described.

RESULTS

In the semi-implantable clinical trials, METF(MET) was -10 to -15 dB for frequencies above 1 kHz, and dropped to -15 to -20 dB for lower frequencies. Fully implantable clinical trial results were 10-20 dB better than the semi-implantable trial results, with less variability at all frequencies. These findings indicate more consistent and better coupling efficiency of the transducer to the ossicles.

摘要

结论

在手术过程中使用最近推出的术中加载器械可提高完全植入式听力设备与听小骨的耦合效率和一致性,并最终使患者使用该设备时表现更佳。

背景

植入式听力设备的患者表现取决于植入物与中耳听小骨的有效耦合。新型术中器械有助于外科医生将植入式听力设备进行耦合,并持续优化患者表现。

材料与方法

Otologics MET听骨刺激器(METF(MET))的中耳传递函数是换能器耦合效率的一种度量,定义为听阈(以dB HL为单位)减去植入阈(以dB MET为单位)。该传递函数在Otologics美国一期半植入式临床试验(未使用术中加载器械)和Otologics美国一期完全植入式临床试验(使用术中加载器械)中进行了测量。两项研究均采用相同的手术技术和换能器设计。文中描述了使用该器械的工作原理和手术技术。

结果

在半植入式临床试验中,对于频率高于1 kHz的情况,METF(MET)为-10至-15 dB,而对于较低频率则降至-15至-20 dB。完全植入式临床试验结果比半植入式试验结果好10 - 20 dB,且在所有频率下变异性更小。这些发现表明换能器与听小骨的耦合效率更一致且更佳。