Accuracy and feasibility of using an electrogoniometer for measuring simple thumb movements.

作者信息

Jonsson Per, Johnson Peter W, Hagberg Mats

机构信息

Occupational and Environmental Medicine, Sahlgrenska University Hospital and Academy at Göteborg University, Göteborg, Sweden.

出版信息

Ergonomics. 2007 May;50(5):647-59. doi: 10.1080/00140130601164490.

Abstract

The aim of this study was to determine the accuracy and feasibility of using an electrogoniometer (Model SG 110; Biometrics, Gwent, UK) for measuring simple thumb movements. Thumb disorders have been associated with the use of hand held devices such as mobile phones and these devices have become an integral part of modern life. In 15 young subjects, the measurements of eight flexion/extension (Flex/Ext) and adduction/abduction (Ad/Ab) thumb positions were compared between a thumb-mounted electrogoniometer and manual goniometer (which was taken as the benchmark). Group mean electrogoniometric measurement errors were below 4 degrees and 5 degrees for Ad/Ab and Flex/Ext measurements, respectively. During mobile phone use, the electrogoniometers measured differences in maximal joint angle postures, which appeared to be related to differences in mobile phone size. High movement velocities may increase the risk of musculoskeletal injury and the results indicated that Ad/Ab movements were twice the speed of Flex/Ext movements during mobile phone use. Electrogoniometers have utility for studying thumb movements during mobile phone use and may be used to evaluate other thumb-based input devices.

摘要

本研究的目的是确定使用电子角度计(型号SG 110;生物测量公司,英国格温特郡)测量简单拇指运动的准确性和可行性。拇指疾病与使用手机等手持设备有关,而这些设备已成为现代生活中不可或缺的一部分。在15名年轻受试者中,将安装在拇指上的电子角度计与手动角度计(作为基准)对八个拇指屈伸(Flex/Ext)和内收/外展(Ad/Ab)位置的测量结果进行了比较。Ad/Ab和Flex/Ext测量的组平均电子角度测量误差分别低于4度和5度。在使用手机期间,电子角度计测量了最大关节角度姿势的差异,这似乎与手机尺寸的差异有关。高运动速度可能会增加肌肉骨骼损伤的风险,结果表明在使用手机期间,Ad/Ab运动速度是Flex/Ext运动速度的两倍。电子角度计在研究使用手机时的拇指运动方面具有实用性,可用于评估其他基于拇指的输入设备。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索