Rosen Yael, Bloch B Nicolas, Lenkinski Robert E, Greenman Robert L, Marquis Robert P, Rofsky Neil M
Department of Radiology, Beth Israel Deaconess Medical Center, Harvard Medical School, Boston, Massachusetts 02215, USA.
Magn Reson Med. 2007 May;57(5):898-904. doi: 10.1002/mrm.21166.
Most prostate MRI/MRS examinations are performed with an endorectal coil inflated with air, leading to an air-tissue interface that induces magnetic susceptibility gradients within the gland. Inflation of the coil with a barium sulfate suspension is described and compared to inflation with air or liquid perfluorocarbon (PFC). The B(0) field in the prostate gland was mapped for five healthy volunteers when the endorectal coil was inflated with each of the three agents. A marked decrease in the posterior-anterior (P-A) field gradient and a significant improvement in field homogeneity were evident in the presence of a barium suspension and PFC relative to air. MRS data acquired from the prostate gland in the presence of air, PFC, and a barium suspension in the endorectal coil showed similar trends, demonstrating improvement in line-widths and spectral resolution when the barium suspension or the PFC were inflating the endorectal coil. On this basis we conclude that a barium suspension provides an available, cheap, and safe alternative to PFC, and we suggest that inflating the endorectal coil with a barium suspension should be considered for prostate MR studies, especially at high field strengths (such as 3T).
大多数前列腺MRI/MRS检查是使用充有空气的直肠内线圈进行的,这会导致气-组织界面,从而在腺体内部诱发磁化率梯度。本文描述了用硫酸钡悬浮液充盈线圈的情况,并将其与用空气或全氟碳化合物(PFC)充盈进行了比较。当用三种介质中的每一种充盈直肠内线圈时,对五名健康志愿者的前列腺内B(0)场进行了测绘。相对于空气,在存在硫酸钡悬浮液和PFC的情况下,前后(P-A)场梯度明显降低,场均匀性显著改善。在直肠内线圈充有空气、PFC和硫酸钡悬浮液的情况下,从前列腺获取的MRS数据显示出类似的趋势,表明当硫酸钡悬浮液或PFC充盈直肠内线圈时,线宽和光谱分辨率有所改善。在此基础上,我们得出结论,硫酸钡悬浮液是PFC的一种可用、廉价且安全的替代品,并且我们建议在前列腺MR研究中,尤其是在高场强(如3T)下,应考虑用硫酸钡悬浮液充盈直肠内线圈。