Lebel R Marc, Wilman Alan H
Department of Biomedical Engineering, University of Alberta, Edmonton, Canada.
Magn Reson Med. 2007 May;57(5):972-5. doi: 10.1002/mrm.21220.
We present a simple and intuitive means for determining the flip angles (FAs) required for smooth transitions between static pseudo steady states (SPSSs) in fast spin echo (FSE) imaging with variable FA (VFA) echo trains. We demonstrate the effectiveness of single and multiple transition pulses to successfully vary refocusing FAs while retaining high signal levels. The graphical interpretation presented here is consistent with previous analytical techniques and permits accurate signal-intensity predictions along the echo train.
我们提出了一种简单直观的方法,用于确定在具有可变翻转角(VFA)回波链的快速自旋回波(FSE)成像中,静态伪稳态(SPSS)之间平滑过渡所需的翻转角(FA)。我们展示了单过渡脉冲和多过渡脉冲在成功改变重聚焦FA的同时保持高信号水平的有效性。这里给出的图形解释与先前的分析技术一致,并允许沿回波链进行准确的信号强度预测。