评估静脉输液调节装置滴注速率准确性的方法。

Method for evaluating drip-rate accuracy of intravenous flow-regulating devices.

作者信息

Carleton B C, Cipolle R J, Larson S D, Canafax D M

机构信息

Department of Pharmacy Practice, College of Pharmacy, University of Minnesota (UM), Minneapolis 55455.

出版信息

Am J Hosp Pharm. 1991 Nov;48(11):2422-6.

PMID:1746576
Abstract

The effects of patient movement and position on the drip-rate accuracy of several i.v. flow-regulating devices were investigated. Intravenous infusion sites were established in 20 healthy adult volunteers. All the subjects received 5% dextrose injection through the same type of i.v. tubing from 500-mL bags hung from standard i.v. fluid poles. The flow-regulating devices tested were the IVAC 280, which served as the control device; a roller clamp; the Dial-A-Flo; the Exacdrop; and the 3M IV Flow Regulator. Drip rates were present at 40 drops/min and were measured before and after the subjects moved among the supine, sitting, and standing positions and walking. The drip rate was reset to 40 drops/min after each position change. Changing position from supine to sitting did not affect mean drip rates for the IVAC 280 and 3M IV Flow Regulator devices but significantly decreased the rates for the roller clamp, Dial-A-Flo, and Exacdrop. The change from sitting to standing did not affect the IVAC 280 and 3M IV Flow Regulator drip rates but significantly decreased the rates for the other devices. None of the rates was dramatically affected when the subjects went from standing to walking, although the effect achieved significance for the roller clamp and Exacdrop devices. The change from walking to the supine position did not affect the drip rates for the IVAC 280 and 3M IV Flow Regulator but significantly increased the rates for the other devices. The drip-rate accuracy of the roller clamp, Dial-A-Flo, and Exacdrop devices was significantly affected when subjects changed positions.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

研究了患者移动和体位对几种静脉输液流量调节装置滴注速率准确性的影响。在20名健康成年志愿者身上建立了静脉输液部位。所有受试者通过连接在标准静脉输液杆上的500毫升袋装同类型静脉输液管接受5%葡萄糖注射液。所测试的流量调节装置包括作为对照装置的IVAC 280;一个滚轮夹;Dial-A-Flo;Exacdrop;以及3M静脉输液流量调节器。滴注速率设定为40滴/分钟,并在受试者在仰卧、坐姿、站立位之间移动以及行走前后进行测量。每次体位改变后,滴注速率重新设定为40滴/分钟。从仰卧位变为坐姿不影响IVAC 280和3M静脉输液流量调节器装置的平均滴注速率,但显著降低了滚轮夹、Dial-A-Flo和Exacdrop的速率。从坐姿变为站立位不影响IVAC 280和3M静脉输液流量调节器的滴注速率,但显著降低了其他装置的速率。当受试者从站立位变为行走时,没有任何一种速率受到显著影响,尽管滚轮夹和Exacdrop装置的影响达到了显著水平。从行走变为仰卧位不影响IVAC 280和3M静脉输液流量调节器的滴注速率,但显著提高了其他装置的速率。当受试者改变体位时,滚轮夹、Dial-A-Flo和Exacdrop装置的滴注速率准确性受到显著影响。(摘要截短于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验