Smith Clara R, DePrince Randolph B, Dackor Jennifer, Weigl Debra, Griffith Jack, Persmark Magnus
BioPharmaceutical Product Development-US, GlaxoSmithKline, 5 Moore Drive, Research Triangle Park, NC 27709, USA.
J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Jul 1;854(1-2):121-7. doi: 10.1016/j.jchromb.2007.04.005. Epub 2007 Apr 13.
We sought to establish a single anion-exchange HPLC method for the separation of linear, open circular and supercoiled plasmid topoisomers using purified topoisomeric forms of three plasmids (3.0, 5.5 and 7.6 kb). However, finding one condition proved elusive as the topoisomer elution order was determined to depend on salt gradient slope. The observed change in selectivity increased with plasmid size and was most pronounced for the linear form. Indeed, the elution order of the linear 7.6 kb plasmid was reversed relative to the supercoiled form. This observation may have implications for methods used in quality control of plasmid DNA.