New J-3 flexible-leaflet polyurethane heart valve prosthesis with improved hydrodynamic performance.

作者信息

Jansen J, Willeke S, Reiners B, Harbott P, Reul H, Rau G

机构信息

Helmholtz-Institute for Biochemical Engineering, Aachen, Germany.

出版信息

Int J Artif Organs. 1991 Oct;14(10):655-60.

PMID:1748534
Abstract

The aim of this new three-leaflet valve development was to design a leaflet with minimum membrane stresses during performance. This is achieved by manufacturing the valve leaflets shaped almost flat in a medium opening position. Thus, the leaflets have two stable positions, one with maximum opening area and the other with favorable stress distribution in the closed position. The transition between the two end positions is achieved through a two-dimensional rolling motion without buckling and with minimum membrane stresses (bulge forces). The manufacturing technique is dip-coating in polyurethane. Hydrodynamic evaluation of the J-3 valve in steady and pulsatile flow showed minimum pressure drop compared to other commercially available valves. Laser-Doppler-anemometry studies indicated very low shear stresses in the flow field downstream of the valve. In durability tests prototypes have reached lifetimes of up to 17 years. In conclusion, the J-3 valve shows superior hydrodynamic performance thereby reducing potential thrombus formation. Minimization of stresses within the valve leaflets through design could reduce calcification.