He Wang, Qin Xu, Jiejing Ren, Gengying Li
Shanghai Key Laboratory of Functional Magnetic Resonance Imaging, East China Normal University, Shanghai 200062, People's Republic of China.
Rev Sci Instrum. 2007 Jan;78(1):015102. doi: 10.1063/1.2424426.
An alternative technique that uses frequency domain multiplexing to acquire phased array magnetic resonance images is discussed in detail. The proposed method has advantages over traditional independent receiver chains in that it utilizes an analog-to-digital converter and a single-chip multicarrier receiver with high performance to reduce the size and cost of the phased array receiver system. A practical four-channel digital receiver using frequency domain multiplexing was implemented and verified on a home-built 0.3 T magnetic resonance imaging system. The experimental results confirmed that the cross talk between each channel was below -60 dB, the phase fluctuations were about 1 degrees , and there was no obvious signal-to-noise ratio degradation. It is demonstrated that the frequency domain multiplexing is a valuable and economical technique, particularly for array coil systems where the multichannel receiver is indispensable and dynamic range is not a critical problem.
详细讨论了一种使用频域复用获取相控阵磁共振图像的替代技术。所提出的方法相对于传统的独立接收器链具有优势,因为它利用了高性能的模数转换器和单芯片多载波接收器来减小相控阵接收器系统的尺寸和成本。在一个自制的0.3T磁共振成像系统上实现并验证了一个使用频域复用的实用四通道数字接收器。实验结果证实,每个通道之间的串扰低于-60dB,相位波动约为1度,并且没有明显的信噪比下降。结果表明,频域复用是一种有价值且经济的技术,特别是对于多通道接收器不可或缺且动态范围不是关键问题的阵列线圈系统。