Porter J R, Wright S M, Famili N
Department of Electrical Engineering, Texas A&M University, College Station 77843-3128.
Magn Reson Med. 1994 Oct;32(4):499-504. doi: 10.1002/mrm.1910320412.
The concept of simultaneous image acquisition as originally conceived by Hyde has been implemented by Roemer et al. using a multiple receiver system. This article describes an alternative technique that uses time domain multiplexing (TDM) to acquire simultaneous images using a single receiver channel. This method requires few modifications to the existing receiver and has been shown to be equivalent to a true four channel receiver in most applications. The multiplexing technique was implemented and tested on a standard commercial scanner that has also been equipped with a four channel receiver. Signal-to-noise results demonstrate that four independent images can be multiplexed through a single receiver channel with no degradation in image quality. Additionally, spine images that were obtained from a normal volunteer with both a multiplexed single channel receiver and a true four channel receiver system are presented.