The aim of the work is to quantitatively compare the effect of the energy separation in the k-edge digital subtraction imaging technique. Images of a custom-made, iodine filled (k-edge = 33.17 keV) test object have been acquired with monochromatic x-ray beams (energy spread <0.1 keV) at the ID17 biomedical beamline of the ESRF. Image acquisition has been performed using two energy separations, namely 0.65 keV (32.85 keV and 33.5 keV, respectively) and 4.4 keV (31.2 keV and 35.6 keV, respectively), using beams of energies on either side of the iodine k-edge. Signal and signal-to-noise ratio (SNR) analysis has been performed as a function of DeltaE and the contrast medium concentrations. The results show that the SNR values measured for DeltaE < 1 keV are only slightly higher than those measured for DeltaE = 4.4 keV. This preliminary study shows that monochromaticity and the energy separation obtained with quasi monochromatic beams from conventional x-ray sources might be suitable for this imaging technique.
这项工作的目的是定量比较k边数字减影成像技术中能量分离的效果。在欧洲同步辐射装置(ESRF)的ID17生物医学光束线处,使用单色X射线束(能量展宽<0.1 keV)采集了一个定制的、充满碘(k边=33.17 keV)的测试物体的图像。使用碘k边两侧能量的光束,以0.65 keV(分别为32.85 keV和33.5 keV)和4.4 keV(分别为31.2 keV和35.6 keV)这两种能量分离进行图像采集。已根据ΔE和造影剂浓度进行了信号及信噪比(SNR)分析。结果表明,对于ΔE < 1 keV测量的SNR值仅略高于对于ΔE = 4.4 keV测量的SNR值。这项初步研究表明,来自传统X射线源的准单色光束所获得的单色性和能量分离可能适用于这种成像技术。