Suppr超能文献

A scanning electron microscopy evaluation of microfractures, deformation and separation in EndoSequence and Profile nickel-titanium rotary files using an extracted molar tooth model.

作者信息

Herold Keven S, Johnson Bradford R, Wenckus Christopher S

机构信息

Department of Endodontics, University of Illinois at Chicago, Chicago, Illinois 60612, USA.

出版信息

J Endod. 2007 Jun;33(6):712-4. doi: 10.1016/j.joen.2007.03.014. Epub 2007 Apr 24.

Abstract

The development of microfractures in the EndoSequence nickel-titanium rotary (NTR) file (Brassler USA, Savannah, GA) and Profile NTR file was evaluated by using scanning electron microscopy (SEM). Seventy-three maxillary buccal roots and 53 mandibular mesial roots with an average canal curvature of 37 degrees were randomly assigned to one of three groups and prepared with 21-mm .06 taper NTR files as follows: (1) EndoSequence at 300 rpm, (2) EndoSequence at 600 rpm, and (3) ProFile at 300 rpm. File sizes .45 to .20 were used in a crown-down technique to a master apical file (MAF) size of .35/.06. The MAF instruments were evaluated by SEM before use, after completion of 7 canals, and after completion of 14 canals. All EndoSequence instruments developed microfractures by the seven-canal evaluation. The ProFile instruments showed no microfractures at the 7- or 14-canal evaluations. EndoSequence files separated at a higher rate than ProFile instruments. Under the conditions of this study, unique file design and electropolishing did not inhibit the development of microfractures in EndoSequence NTR.