Weiss D, Horowitz Y, Oster L
Nuclear Research Center-Negev, Beersheva, Israel.
Radiat Prot Dosimetry. 2007;126(1-4):206-9. doi: 10.1093/rpd/ncm043. Epub 2007 May 19.
The TL signal following 5 eV photon excitation of previously irradiated and readout material has been studied as a function of ionisation density and various experimental parameters: (i) maximum temperature of the first readout; (ii) photon fluence; (iii) photon energy and (iv) beta ray dose. Following alpha particle irradiation, the ratio of the second-readout to first-readout TL signal, epsilon(alpha,) has been found to be 10-20 times higher than that following beta irradiation, indicative of the possibility of using the double ratio epsilon(alpha)/epsilon(beta) as a mixed-field discriminator. The beginning of an attempt to explain this unusual effect is offered in the framework of the track structure theory and kinetic modelling of the beta ray dose-response of the first and second readouts.