Development, validation and application of a chemiluminescent immunoassay for the measurement of circulating chorionic gonadotropin levels in the laboratory macaque.

作者信息

Lohstroh Pete, Laughlin Lisa, Gee Nancy, Lasley Bill

机构信息

Center for Health and the Environment, University of California, Davis, CA 95616-8615, USA.

出版信息

J Med Primatol. 2007 Jun;36(3):164-9. doi: 10.1111/j.1600-0684.2006.00189.x.

Abstract

BACKGROUND

Methods for the measurement of chorionic gonadotropin for pregnancy detection in laboratory macaques chorionic gonadotropin (mCG) have been limited by the paucity of specific reagents. Current assays use reagents developed for human chorionic gonadotropin (hCG) assays and have been limited to radio- or enzyme-immuno assays.

METHODS

This report describes an automated assay for mCG using reagents for hCG detection that has been adapted to the automated chemiluminescence platform of the Bayer ACS-180 autoanalyzer.

RESULTS

This automated assay performs similarly to previous assays in terms of sensitivity and specificity but provides additional speed, reliability, and more easily accessible reagents compared with previous formats.

CONCLUSIONS

When implemented broadly, this assay could provide standardization of mCG measurements for both research and colony management purposes.