Pachnis Ioannis, Demosthenous Andreas, Donaldson Nick
Department of Electronic and Electrical Engineering, University College London, Torrington Place, London WC1E 7JE, UK.
IEEE Trans Biomed Eng. 2007 Jun;54(6 Pt 1):1067-74. doi: 10.1109/TBME.2007.891170.
In this paper, we present a simple passive technique for removing myoelectric interference in neural recording systems using tripolar electrodes. Imbalance is simply unavoidable with the conventional quasi-tripole (QT) amplifier and this technakshnique is based on a modified version of the QT, which sees the tripole as a bridge that can be balanced by adding impedance to one of the outer electrodes. We show that parallel resistance and capacitance is better than a series combination for use at all frequencies, and that with a tripole that was intentionally made imbalanced, by an amount that we measure as 3%, the interference can be reduced about 10-fold. It is important to null the interference at low frequencies, in the electromyography (EMG) band, to get the best improvement. Common-mode interference can also be reduced by appropriate trimming if necessary.
在本文中,我们提出了一种简单的无源技术,用于去除使用三极电极的神经记录系统中的肌电干扰。传统的准三极(QT)放大器不可避免地会出现不平衡,而该技术基于QT的改进版本,它将三极视为一座桥,可以通过向其中一个外部电极添加阻抗来实现平衡。我们表明,在所有频率下,并联电阻和电容比串联组合更适用,并且对于一个故意制造出3%不平衡量的三极电极,干扰可降低约10倍。在肌电图(EMG)频段的低频处消除干扰对于获得最佳改善效果很重要。如有必要,也可通过适当调整来降低共模干扰。