Abitbol O, Duong M-H, Hoang-Xuan T, Gatinel D, Lachkar Y
Fondation Ophtalmologique Adolphe de Rothschild, Paris.
J Fr Ophtalmol. 2007 May;30(5):503-9. doi: 10.1016/s0181-5512(07)89631-8.
To evaluate the reliability of scanning laser polarimetry with variable corneal compensation after laser-assisted in-situ keratomileusis (LASIK).
Thirty-six eyes of 18 consecutive patients who had LASIK where included in the study. For each eye, one scanning laser polarimetry with variable corneal compensation (GDxVCC) was performed before LASIK (GDx no 1). One month after surgery, two other scanning laser polarimetry procedures were performed: one used the corneal compensation measured preoperatively (GDx no 2) and one used a new corneal compensation, measured postoperatively (GDx no 3). The values measured preoperatively were compared first to the values obtained postoperatively with GDx no 2, then to the values obtained postoperatively with GDx no 3 (Wilcoxon test).
When corneal compensation was re-measured postoperatively (GDx no 3), the measurements were reliable for 35 eyes (97%). We found no significant differences between the preoperative and postoperative values except for the NFI (p=0.032). When the preoperative corneal compensation value was used in the postoperative measurements (GDx no 2), the exam was not reliable for 13 eyes, and it could not be performed at all for two eyes. For the remaining eyes, the values of four parameters were significantly modified: superior thickness (p=0.03), superior ratio (p=0.0005), inferior ratio (p=0.009), and ellipse modulation (p=0.39). The values of the other parameters (average thickness and inferior average) remained unchanged.
The use of a preoperative customized corneal compensation for the realization of post-LASIK measurements does not provide reliable results. The GDxTMVCC can be considered a reliable tool after LASIK only when the corneal compensation is re-evaluated postoperatively.
评估准分子原位角膜磨镶术(LASIK)后可变角膜补偿的扫描激光偏振仪的可靠性。
连续18例接受LASIK手术的患者的36只眼纳入本研究。每只眼在LASIK术前进行一次可变角膜补偿的扫描激光偏振仪检查(GDxVCC,第1次GDx)。术后1个月,进行另外两次扫描激光偏振仪检查:一次使用术前测量的角膜补偿值(第2次GDx),另一次使用术后测量的新角膜补偿值(第3次GDx)。术前测量值首先与第2次GDx术后获得的值进行比较,然后与第3次GDx术后获得的值进行比较(Wilcoxon检验)。
术后重新测量角膜补偿值(第3次GDx)时,35只眼(97%)的测量结果可靠。除神经纤维指数外,术前和术后测量值无显著差异(p=0.032)。当术后测量使用术前角膜补偿值时(第2次GDx),13只眼的检查结果不可靠,2只眼根本无法进行检查。对于其余的眼睛,四个参数的值有显著改变:上方厚度(p=0.03)、上方比率(p=0.0005)、下方比率(p=0.009)和椭圆调制(p=0.39)。其他参数(平均厚度和下方平均厚度)的值保持不变。
LASIK术后测量采用术前定制的角膜补偿不能提供可靠的结果。仅当术后重新评估角膜补偿时,GDxTMVCC才可被视为LASIK术后可靠的工具。