Brooks F D, Buffler A, Allie M S, Herbert M S, Nchodu M R, Jones D T L, Smit F D, Nolte R, Dangendorf V
Department of Physics, University of Cape Town, Rondebosch, Cape 7700, South Africa.
Radiat Prot Dosimetry. 2007;126(1-4):218-22. doi: 10.1093/rpd/ncm045. Epub 2007 Jun 16.
A compact liquid organic neutron spectrometer based on a single NE213 liquid scintillator (5 cm diameter x 5 cm) is described. The spectrometer is designed to measure neutron fluence spectra over the energy range 2-200 MeV and is suitable for use in neutron fields having any type of time structure. Neutron fluence spectra are obtained from measurements of two-parameter distributions (counts versus pulse-height and pulse shape) using the Bayesian unfolding code MAXED. Calibration and test measurements made using a pulsed neutron beam with a continuous energy spectrum are described and the application of the spectrometer to radiation dose measurements is discussed.