Sorom Abraham J, Driscoll Colin L W, Beatty Charles W, Lundy Larry
Department of Otolaryngology-Head and Neck Surgery, Mayo Clinic, Rochester, MN 55905, USA.
Otolaryngol Head Neck Surg. 2007 Jul;137(1):65-9. doi: 10.1016/j.otohns.2007.01.007.
To review hearing results after implantation of a self-crimping stapes prosthesis.
Analysis of hearing results in patients implanted with a self-crimping stapes prosthesis at two academic hospitals from 2000 to 2004.
Seventy-nine ears were divided into short-term and intermediate follow-up groups. The mean postoperative air-bone gap (ABG), preoperative minus postoperative ABG, and preoperative minus postoperative bone conduction values were 5.7, 21.4, and 4.1 dB, respectively, for the short-term group, and 6.3, 22.3, and 4.7 dB, respectively, for the intermediate group. The ABG was <10 dB in 88 percent of the short-term group and in 79 percent of the intermediate group.
The self-crimping Nitinol stapes prosthesis provides excellent short-term and intermediate postoperative hearing results, and may overcome the limitations of stapes prostheses requiring manual crimping.
This paper provides evidence for the use of a self-crimping Nitinol stapes prosthesis, which may simplify hearing restoration surgery for stapes fixation.