Huguet J, Bonnin M R, Guillén E, Navarro M A
Sección de Hormonas (Servicio de Bioquímica), Hospital Princeps d'Espanya, Barcelona, España.
Eur J Clin Chem Clin Biochem. 1991 Sep;29(9):565-7.
The analytical performance of the Stratus (Baxter) automatic analyser for human lutropin and follitropin determination was evaluated and compared with that of an immunoradiometric assay. Within-run and between-run imprecision were lower than those of the immunoradiometric assay. No significant differences were obtained in the parallelism study. A good and significant correlation was obtained between both methods. Results obtained for both methods were not interchangeable. The Stratus automatic analyser can replace radioisotopic methods, thus eliminating radioactive hazards.