Suppr超能文献

利用电荷注入器件检测带电粒子。

Detection of charged particles with charge injection devices.

作者信息

Fletcher Kurtis A, Apker Benjamin, Hammond Samantha, Punaro John, Marshall Frederic J, Laine Jonathan, Forties Robert

机构信息

Department of Physics and Astronomy, SUNY Geneseo, 1 College Circle, Geneseo, NY 14454, USA.

出版信息

Rev Sci Instrum. 2007 Jun;78(6):063301. doi: 10.1063/1.2745234.

Abstract

A method for using charge injection devices (CIDs) for detection of high-energy charged particles from inertial-confinement fusion reactions is described. Because of the relatively small depletion region of the CID camera (depletion depth of approximately 7 mum), aluminum foils are placed in front of the device to reduce the energy of the charged particles and maximize the energy deposited in the CID. Simultaneous measurements of (2)H(d,p)(3)H protons with a CID and a surface barrier detector indicate that the CID is an efficient detector of charged fusion products. Tests using high energy alpha particles emitted from a radium-226 source are also reported.

摘要

描述了一种使用电荷注入器件(CID)来探测惯性约束聚变反应中高能带电粒子的方法。由于CID相机的耗尽区相对较小(耗尽深度约为7微米),因此在该器件前放置铝箔以降低带电粒子的能量,并使沉积在CID中的能量最大化。用CID和表面势垒探测器对(2)H(d,p)(3)H质子进行的同步测量表明,CID是一种探测带电聚变产物的有效探测器。还报告了使用镭-226源发射的高能α粒子进行的测试。