Petitjean C, Mairy Y, Saletes M, Fauvel M, Brudon J R
Chirurgien Vasculaire, Clinique Labrouste, Paris.
J Chir (Paris). 1991 Nov;128(11):494-7.
Endocaval filters are often used to prevent pulmonary embolism but they have a number of disadvantages. The DIL filter, made of a memory metal wire, is intended to male up for some of these disadvantages. It acts by modifying the shape of the inferior vena cava, which it filters through its meshed loops. It is inserted percutaneously, causes little trauma, and its release is progressive. However, it requires measuring the caliber of the inferior vena cava. This filter was inserted in thirty-four patients over a period of 13 months. One filter has migrated. No recurrence of pulmonary embolism and no thrombosis of the inferior vena cave occurred.
腔静脉滤器常用于预防肺栓塞,但它们有许多缺点。由记忆金属丝制成的DIL滤器旨在弥补其中一些缺点。它通过改变下腔静脉的形状来发挥作用,它通过其网状环对下腔静脉进行过滤。它通过经皮插入,造成的创伤小,且释放过程是渐进的。然而,它需要测量下腔静脉的管径。在13个月的时间里,该滤器被植入了34名患者体内。有一个滤器发生了移位。没有肺栓塞复发,也没有发生下腔静脉血栓形成。