Katz David, Akiyama Toshio
Department of Internal Medicine, Hospital of the University of Pennsylvania, Philadelphia, PA, USA.
Ann Noninvasive Electrocardiol. 2007 Jul;12(3):223-6. doi: 10.1111/j.1542-474X.2007.00165.x.
To describe VVI-pacemaker longevity by model type at our institution and report on a long-lasting model and the longest-lasting pacemaker to be described in the literature.
Cardiac pacemakers are becoming increasingly common in the United States. Presently their batteries are expected to last up to 12 years. Pacemaker generator change is associated with increased cost to the health care system and is inconvenient for patients.
After identifying a group of very long-lasting CPI Microlith 605 VVI pulse generators, we reviewed records on all patients who had either Guidant or Medtronic pulse generator explantation at our institution over a 10-year period. Average longevities were calculated for all VVI pacemakers, four common VVI models, and the CPI Microlith 605.
A total of 105 VVI-programmed pacemakers were identified. Their average longevity was 7.2 years. The two most common Medtronic VVI-programmed pacemakers explanted were the Thera (7.1 years) and Kappa (7.3 years). The two most common Guidant/CPI models were the Vigor (4.2 years) and Discovery (5.7 years). The CPI Microlith 605 (19.2 years) lasted more than 26 years in one patient before being explanted.
At a time when pacemakers are being used more frequently, pacemaker longevity may decrease as a result of the use of dual-chamber pacing systems. In our study, the CPI Microlith 605 had an average longevity more than twice that of all other VVI pacemakers. We also report on a pulse generator that lasted 26.3 years.
描述我院不同型号VVI起搏器的使用寿命,并报告一种使用寿命长的型号以及文献中所描述的使用寿命最长的起搏器。
心脏起搏器在美国越来越普遍。目前,其电池预计可持续使用长达12年。更换起搏器发生器会增加医疗系统的成本,且给患者带来不便。
在确定一组使用寿命极长的CPI Microlith 605 VVI脉冲发生器后,我们回顾了我院10年间所有接受了Guidant或美敦力脉冲发生器植入的患者的记录。计算了所有VVI起搏器、四种常见VVI型号以及CPI Microlith 605的平均使用寿命。
共识别出105个程控VVI起搏器。它们的平均使用寿命为7.2年。被植入的两种最常见的美敦力程控VVI起搏器是Thera(7.1年)和Kappa(7.3年)。两种最常见的Guidant/CPI型号是Vigor(4.2年)和Discovery(5.7年)。CPI Microlith 605(19.2年)在一名患者体内使用了超过26年才被取出。
在起搏器使用越来越频繁的时代,由于双腔起搏系统的使用,起搏器的使用寿命可能会降低。在我们的研究中,CPI Microlith 605的平均使用寿命是所有其他VVI起搏器的两倍多。我们还报告了一个使用了26.3年的脉冲发生器。