Song Ruitian, Cohen Alan R, Song Hee Kwon
Laboratory for Structural NMR Imaging, Department of Radiology, University of Pennsylvania Medical Center, Philadelphia, Pennsylvania, USA.
J Magn Reson Imaging. 2007 Jul;26(1):208-14. doi: 10.1002/jmri.20994.
To develop and validate an optimized respiratory-gated, gradient-echo sampling of free induction decay and echo (GESFIDE) pulse sequence for the simultaneous measurement of R2, R2*, and R2' in the liver or heart.
Fifteen subjects (12 thalassemia patients and three normal volunteers) were scanned using an optimized navigator-gated GESFIDE pulse sequence for the measurement of R2, R2*, and R2' in the liver and heart. For imaging the myocardium, dark-blood preparation was used to suppress the blood signal to improve accuracy. The results were compared with those obtained from breath-held GESFIDE and multi-gradient-echo (GRE) scans.
Good agreement between breath-held and navigator-gated scans was found for R2, R2*, and R2' values in the liver (slopes = 0.97-0.99, r = 0.997-0.998, P < 0.0001) and for R2* in the heart (slope = 1.02, r = 0.85, P < 0.0001). Both R2* and R2' were closely correlated to R2 in the liver, with correlation factors of 0.998 and 0.994, respectively, but weaker correlations were observed in the heart (r = 0.72 for R2* vs. R2 and r = 0.51 for R2' vs. R2).
The improved sequence enables free-breathing measurements of transverse relaxation rates of the myocardium and liver. The method precludes the need for multiple breath-held scans and possible misregistration issues, and may prove most beneficial for imaging young children and patients who may have difficulty with prolonged or repeated breath-holds.
开发并验证一种优化的呼吸门控、自由感应衰减与回波梯度回波采样(GESFIDE)脉冲序列,用于同时测量肝脏或心脏中的R2、R2*和R2'。
使用优化的导航门控GESFIDE脉冲序列对15名受试者(12名地中海贫血患者和3名正常志愿者)进行扫描,以测量肝脏和心脏中的R2、R2*和R2'。为了对心肌成像,采用黑血准备技术抑制血液信号以提高准确性。将结果与屏气GESFIDE和多梯度回波(GRE)扫描获得的结果进行比较。
在肝脏中,屏气扫描和导航门控扫描的R2、R2和R2'值之间具有良好的一致性(斜率=0.97 - 0.99,r = 0.997 - 0.998,P < 0.0001),在心脏中R2的一致性也较好(斜率=1.02,r = 0.85,P < 0.0001)。肝脏中R2和R2'均与R2密切相关,相关系数分别为0.998和0.994,但在心脏中相关性较弱(R2与R2的r = 0.72,R2'与R2的r = 0.51)。
改进后的序列能够在自由呼吸状态下测量心肌和肝脏的横向弛豫率。该方法无需多次屏气扫描以及可能出现的配准错误问题,对于幼儿和难以长时间或重复屏气的患者成像可能最为有益。