用于同时测量可逆(R2')和不可逆(R2)横向弛豫率的回波间隔优化。

Echo-spacing optimization for the simultaneous measurement of reversible (R2') and irreversible (R2) transverse relaxation rates.

作者信息

Song Ruitian, Song Hee Kwon

机构信息

Laboratory for Structural NMR Imaging, Department of Radiology, University of Pennsylvania Medical Center, Philadelphia, PA 19104, USA.

出版信息

Magn Reson Imaging. 2007 Jan;25(1):63-8. doi: 10.1016/j.mri.2006.09.008. Epub 2006 Nov 13.

Abstract

Accurate measurement of reversible (R2') and irreversible (R2) transverse relaxation rates plays a key role in various magnetic resonance imaging research and applications. Although optimization of echo spacing for a multiecho pulse sequence measuring a single exponential decay has been investigated, optimization in sequences such as Gradient-Echo Sampling of Free Induction Decay and Echo (GESFIDE), in which two echo trains are simultaneously measured to obtain both R2 and R2', has not been reported. In this work, optimum echo spacings for the GESFIDE sequence are determined to improve the accuracy of measured relaxation parameters. Various relaxation rates and the number of acquired echoes are considered, as well as whether the receiver bandwidth is kept fixed or is varied with echo spacing. In the case of constant receiver bandwidth, results show that the echo train length approximately equal to T2* should be used for each echo train in GESFIDE to minimize uncertainty in R2 or R2'. If the receiver bandwidth is allowed to change with echo spacing in order to maximize the image signal-to-noise ratio, the optimum echo train length will vary, generally increasing with the number of echoes.

摘要

准确测量可逆横向弛豫率(R2')和不可逆横向弛豫率(R2)在各种磁共振成像研究及应用中起着关键作用。尽管针对测量单指数衰减的多回波脉冲序列的回波间隔优化已得到研究,但对于诸如自由感应衰减与回波梯度回波采样(GESFIDE)这类同时测量两个回波串以获取R2和R2'的序列,尚未有关于其优化的报道。在本研究中,确定了GESFIDE序列的最佳回波间隔,以提高所测弛豫参数的准确性。考虑了各种弛豫率、采集回波的数量,以及接收带宽是保持固定还是随回波间隔变化的情况。在接收带宽恒定的情况下,结果表明,在GESFIDE中,每个回波串应使用近似等于T2*的回波串长度,以最小化R2或R2'的不确定性。如果允许接收带宽随回波间隔变化以最大化图像信噪比,则最佳回波串长度会有所不同,通常会随着回波数量增加。

相似文献

1
Echo-spacing optimization for the simultaneous measurement of reversible (R2') and irreversible (R2) transverse relaxation rates.
Magn Reson Imaging. 2007 Jan;25(1):63-8. doi: 10.1016/j.mri.2006.09.008. Epub 2006 Nov 13.
2
Reversible, irreversible and effective transverse relaxation rates in normal aging brain at 3T.
Neuroimage. 2014 Jan 1;84:1032-41. doi: 10.1016/j.neuroimage.2013.08.051. Epub 2013 Sep 1.
4
Comparison of R2' measurement methods in the normal brain at 3 Tesla.
Magn Reson Med. 2015 Mar;73(3):1228-36. doi: 10.1002/mrm.25232. Epub 2014 Apr 18.
5
Benchmarking transverse spin relaxation based oxygenation measurements in the brain during hypercapnia and hypoxia.
J Magn Reson Imaging. 2017 Sep;46(3):704-714. doi: 10.1002/jmri.25582. Epub 2017 Mar 17.
6
Field strength and angle dependence of trabecular bone marrow transverse relaxation in the calcaneus.
J Magn Reson Imaging. 1997 Mar-Apr;7(2):382-8. doi: 10.1002/jmri.1880070222.

引用本文的文献

1
Rapid simultaneous estimation of relaxation rates using multi-echo, multi-contrast MRI.
Magn Reson Imaging. 2024 Oct;112:116-127. doi: 10.1016/j.mri.2024.07.007. Epub 2024 Jul 4.
3
GESFIDE-PROPELLER approach for simultaneous R2 and R2* measurements in the abdomen.
Magn Reson Imaging. 2013 Dec;31(10):1760-5. doi: 10.1016/j.mri.2013.08.003. Epub 2013 Sep 14.
4
Iron in chronic brain disorders: imaging and neurotherapeutic implications.
Neurotherapeutics. 2007 Jul;4(3):371-86. doi: 10.1016/j.nurt.2007.05.006.

本文引用的文献

2
Single spin-echo proton transverse relaxometry of iron-loaded liver.
NMR Biomed. 2004 Nov;17(7):446-58. doi: 10.1002/nbm.905.
3
Dose resolution in radiotherapy polymer gel dosimetry: effect of echo spacing in MRI pulse sequence.
Phys Med Biol. 2001 Feb;46(2):449-60. doi: 10.1088/0031-9155/46/2/312.
4
High-speed spectroscopic imaging for cancellous bone marrow R(2)* mapping and lipid quantification.
Magn Reson Imaging. 2000 Sep;18(7):777-86. doi: 10.1016/s0730-725x(00)00165-x.
5
Complications of nonlinear echo time spacing for measurement of T (2).
NMR Biomed. 2000 Jan;13(1):1-7. doi: 10.1002/(sici)1099-1492(200002)13:1<1::aid-nbm603>3.0.co;2-e.
7
Improving calibration accuracy in gel dosimetry.
Phys Med Biol. 1998 Oct;43(10):2709-20. doi: 10.1088/0031-9155/43/10/002.
8
Proximal femur: assessment for osteoporosis with T2* decay characteristics at MR imaging.
Radiology. 1998 Nov;209(2):531-6. doi: 10.1148/radiology.209.2.9807585.
9
Optimal time spacings for T2 measurements: monoexponential and biexponential systems.
NMR Biomed. 1998 Oct;11(6):297-305. doi: 10.1002/(sici)1099-1492(199810)11:6<297::aid-nbm531>3.0.co;2-a.
10
Measurement of magnetic resonance T2 for physiological experiments.
J Appl Physiol (1985). 1997 Sep;83(3):904-11. doi: 10.1152/jappl.1997.83.3.904.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验