Suppr超能文献

使用腹腔镜筋膜闭合装置进行主动脉穿刺。

Aortic puncture with a laparoscopic fascial closure device.

作者信息

Lee Gene, Nguyen Alexander, Kivnick Seth, Marshall John R

机构信息

Department of Obstetrics and Gynecology, Harbor-University of California Los Angeles, Torrance, California, USA.

出版信息

Obstet Gynecol. 2007 Aug;110(2 Pt 2):533-5. doi: 10.1097/01.AOG.0000269143.63568.c6.

Abstract

BACKGROUND

Vascular injury during laparoscopic surgery typically occurs with the blind insertion of either the Veress needle or the trocar. We report an aortic puncture from the use of a laparoscopic fascial closure device.

CASE

After a laparoscopic-assisted vaginal hysterectomy and during closure of the 10-mm umbilical port with a fascial closure device, attempted passage of the needle encountered unusual resistance in the abdominal wall. Application of increased force resulted in uncontrolled entry of the needle into the abdominal cavity and a 1-mm puncture of the aorta. After emergency laparotomy the puncture was successfully repaired by a vascular surgeon.

CONCLUSION

A previously unreported complication of this laparoscopic fascial closure device is aortic injury. Unusual resistance to passage of the needle should engender extra caution. The use of assistive closure devices should be reserved for patients with difficult anatomy.

摘要

背景

腹腔镜手术中的血管损伤通常发生在盲目插入Veress针或套管针时。我们报告了一例因使用腹腔镜筋膜闭合装置导致的主动脉穿刺。

病例

在腹腔镜辅助下阴式子宫切除术后,使用筋膜闭合装置闭合10毫米脐部切口时,进针遇到腹壁异常阻力。加大力度后,针失控进入腹腔,导致主动脉出现1毫米的穿刺伤。紧急剖腹手术后,血管外科医生成功修复了穿刺伤。

结论

这种腹腔镜筋膜闭合装置此前未被报道的并发症是主动脉损伤。进针遇到异常阻力时应格外小心。辅助闭合装置应仅用于解剖结构复杂的患者。