Yokoyama T, Kinoshita T
Central Research Laboratory, SS Pharmaceutical Co. Ltd., Chiba, Japan.
J Chromatogr. 1991 Jul 17;568(1):212-8. doi: 10.1016/0378-4347(91)80355-g.
A high-performance liquid chromatographic method is described for the selective determination of taurine in biological fluids by post-column fluorescence reaction. Taurine was separated on an adsorption-distribution type Shodex Ionpac KC-811 column. Then it was converted with hypochlorite into the corresponding N-chloramine, which was allowed to react with thiamine to give fluorescent thiochrome. As little as 6 ng per injection of taurine could be determined. The average recoveries of spiked taurine in serum and urine were 99.5 +/- 2.7 and 101.8 +/- 2.9%, respectively. The method could be applied to the assay of taurine in human serum and urine with simple pretreatment.