Dini Sharifeh A, Awan Shahid B, Dou Kai, Meigooni Ali S
Department of Radiation Medicine, University of Kentucky Chandler Medical Center, Lexington, Kentucky, U.S.A.
J Appl Clin Med Phys. 2007 Jul 17;8(3):60-75. doi: 10.1120/jacmp.v8i3.2435.
Recently, to eliminate problems associated with seed type sources, RadioMed Corporation (Tyngsboro, MA) introduced RadioCoil 103Pd sources for interstitial prostate implants. The RadioCoil sources are available in integral lengths ranging from 1.0 cm to 6.0 cm. In this project, dosimetric characteristics of these sources were determined following the TG-43U1 recommendations, with consideration of our recent publication on the evaluation of two-dimensional anisotropy function for elongated brachytherapy sources. Dosimetric parameters of these sources were determined experimentally in Solid Water (Gammex RMI, Middleton, WI) and theoretically using Monte Carlo simulation in Solid Water and liquid water. Per the TG-43U1 protocol, the consensus of these results would be used for their clinical applications.