Measuring quaternary ammonium cleaning agents with ion selective electrodes.

作者信息

Plesha Michael A, Van Wie Bernard J, Mullin James M, Kidwell David A

机构信息

Chemical Engineering Department, Washington State University, P.O. Box 2710 Pullman, WA 99164-2710, USA.

出版信息

Anal Chim Acta. 2006 Jun 16;570(2):186-94. doi: 10.1016/j.aca.2006.04.024. Epub 2006 Jun 2.

Abstract

Data for coated-wire, ion selective electrodes (ISEs) are presented for cationic surfactant ions found in common cleaners including benzyldimethyltetradecylammonium, benzyldimethyldodecylammonium, and benzyldimethylhexadecylammonium. The ion exchangers dinonylnaphthalene sulfonic acid, tetraphenyborate, and tetrakis(4-chlorophenyl)borate are examined, showing dinonylnaphthalene sulfonic acid to be the favored species. The ISEs exhibit approximately Nernstian behavior down to the 10(-6) M limit of detection with lifetimes in excess of 50 days when used continuously, and a shelf life of over 100 days. Reaching the upper detection limit at the critical micelle concentration requires use of polymeric-membrane reference electrodes including a new membrane cocktail, which allow response measurements of an order of magnitude higher than the traditional fritted-glass reference electrode. The surfactant ISEs show excellent selectivity over the common metal ions Na+, K+, Mg2+, Ca2+, and Cu2+ with selectivity coefficients less than 10(-5.3). The ISEs are also selective over the lower molecular weight quaternary ammonium ions tetradecyltrimethylammonium, dodecyltrimethylammonium, benzyldimethyl(2-hydroxyethyl)ammonium, and tetrabutylammonium with selectivity coefficients ranging from 10(-1.7) to 10(-5.5). Use of a single electrode to determine accurately the total cationic surfactant concentration in common cleaning solutions is accomplished with information about concentration dependent interferences and a modified Nikolsky-Eisenman model. Finally, quaternary ammonium surfactants have a deleterious effect on the measurements of pH and common ions like K+, Mg2+ and Ca2+ with polymeric ISEs. This makes it critical to include surfactant electrodes in a detector array when cleaning agents are present.

摘要

本文给出了用于检测常见清洁剂中阳离子表面活性剂离子(包括苄基二甲基十四烷基铵、苄基二甲基十二烷基铵和苄基二甲基十六烷基铵)的涂丝离子选择性电极(ISE)的数据。对离子交换剂二壬基萘磺酸、四苯硼酸盐和四(4-氯苯基)硼酸盐进行了研究,结果表明二壬基萘磺酸是最适宜的物质。这些ISE在低至10⁻⁶ M的检测限下表现出近似能斯特行为,连续使用时寿命超过50天,保质期超过100天。在临界胶束浓度达到检测上限时,需要使用包括新型膜混合液的聚合物膜参比电极,其响应测量值比传统烧结玻璃参比电极高一个数量级。表面活性剂ISE对常见金属离子Na⁺、K⁺、Mg²⁺、Ca²⁺和Cu²⁺具有出色的选择性,选择性系数小于10⁻⁵.³。ISE对较低分子量的季铵离子十四烷基三甲基铵、十二烷基三甲基铵、苄基二甲基(2-羟乙基)铵和四丁基铵也具有选择性,选择性系数在10⁻¹.⁷至10⁻⁵.⁵之间。利用浓度依赖性干扰信息和改进的尼科尔斯基-艾森曼模型,使用单个电极可准确测定常见清洁溶液中的总阳离子表面活性剂浓度。最后,季铵表面活性剂对聚合物ISE测量pH值以及K⁺、Mg²⁺和Ca²⁺等常见离子有有害影响。因此,当存在清洁剂时,在检测器阵列中包含表面活性剂电极至关重要。

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验