Quality assurance of brachytherapy afterloaders using a multi-slit phantom.

作者信息

Kohr Peter, Siebert Frank-André

机构信息

University Clinic of Schleswig-Holstein, Campus Kiel, Clinic of Radiotherapy, Arnold-Heller-Str. 9, 24105 Kiel, Germany.

出版信息

Phys Med Biol. 2007 Sep 7;52(17):N387-91. doi: 10.1088/0031-9155/52/17/N03. Epub 2007 Aug 16.

Abstract

Brachytherapy is a very successful treatment option for many different tumors. Consequently the quality assurance (QA) of afterloaders is of special interest. For many quality checks verification films are essential. Because currently in many clinics the film development units are replaced by computed radiography systems, the question arises of how to perform QA for afterloaders without conventional radiography films. In this note the use of a new designed multi-slit phantom demonstrates the possibility of performing QA checks with high sensitivity computed radiography systems. Because the presented phantom allows a method for imaging the radiation source it might be useful in the development of further QA techniques.

摘要

近距离放射治疗是许多不同肿瘤非常成功的治疗选择。因此,后装治疗机的质量保证(QA)备受关注。对于许多质量检查而言,验证片至关重要。由于目前许多诊所的胶片冲洗设备已被计算机放射成像系统取代,于是出现了如何在没有传统放射胶片的情况下对后装治疗机进行质量保证的问题。在本说明中,一种新设计的多狭缝模体的使用证明了使用高灵敏度计算机放射成像系统进行质量保证检查的可能性。由于所展示的模体提供了一种对辐射源进行成像的方法,它可能对进一步质量保证技术的开发有用。