Repeatability of dynamic eye pupil response measurement using the Pupilscan instrument.

作者信息

Murray A, Lawrence G P, Clayton R H

机构信息

Regional Medical Physics Department, Freeman Hospital, Newcastle upon Tyne, UK.

出版信息

Clin Phys Physiol Meas. 1991 Nov;12(4):377-85.

PMID:1778037
Abstract

The repeatability of pupil response measurements made with the Pupilscan instrument has been assessed. Two operators made 32 repeat recordings of the pupil response in each eye of eight normal subjects over four days. A total of 512 recordings were made and the results analysed by variance analysis. Room light levels were carefully controlled. The instrument measured initial pupil diameter, and after a 0.2 s light stimulus measured minimum and final diameters, maximum diameter change, constriction velocity and time to minimum diameter. For all measurements except time to minimum, there were significant differences (P less than 0.001) between the normal subjects. Day-to-day variation was also significant (P less than 0.005), but did not mask the differences between subjects. It is concluded that the average of at least four repeat measurements are made, giving error ranges similar to those observed from day-to-day changes.

摘要

已对使用瞳孔扫描仪仪器进行的瞳孔反应测量的可重复性进行了评估。两名操作人员在四天内对八名正常受试者的每只眼睛进行了32次瞳孔反应重复记录。总共进行了512次记录,并通过方差分析对结果进行了分析。室内光照水平得到了严格控制。该仪器测量初始瞳孔直径,在0.2秒的光刺激后测量最小和最终直径、最大直径变化、收缩速度以及达到最小直径的时间。除达到最小直径的时间外,所有测量在正常受试者之间均存在显著差异(P小于0.001)。每日变化也很显著(P小于0.005),但并未掩盖受试者之间的差异。得出的结论是,至少进行四次重复测量的平均值,其误差范围与每日变化所观察到的相似。