Suppr超能文献

[麻醉期间无创连续血压监测的动脉张力测量法临床评估]

[Clinical evaluation of arterial tonometry for noninvasive, continuous blood pressure monitoring during anesthesia].

作者信息

Ueda M

机构信息

Department of Anesthesiology, Hokkaido University School of Medicine, Sapporo, Japan.

出版信息

Hokkaido Igaku Zasshi. 1991 Nov;66(6):805-17.

PMID:1783369
Abstract

A new continuous, non-invasive arterial blood pressure monitoring system based on arterial tonometry was currently developed. A sensor of this system consists of an array of 15 piezoresistive pressure transducers. An automatic sensor positioning system ensure that at least one of the 15 transducers is positioned on the center of the pneumatically flattened, but not occluded, portion of the artery. Accuracy, reliability and clinical acceptance of this system were evaluated by comparing tonometric blood pressure (TBP) measurements with intra-arterial blood pressure (IBP) measurements in 70 normotensive anesthetized patients (normotensive group) and 21 hypotensive anesthetized patients (hypotensive group). IBP was measured in either the right or left radial artery by a Gould P23XL calibrated transducer while TBP was measured at the radial artery of the other side. The outputs of the two systems were recorded on a PCM tape recorder for later analysis. The TBP waveform was similar to the IBP waveform. A significantly high correlation existed between the two methods for systolic, mean and diastolic measurements. Magnitudes of bias (mean difference between the two methods) of the normotensive group were -0.4 +/- 5.9 (mean +/- SD) mmHg for systolic, -0.1 +/- 4.7 mmHg for mean and 0.0 +/- 5.4 mmHg for diastolic measurements and those of the hypotensive group were -0.7 +/- 6.6 mmHg for systolic, 0.7 +/- 5.0 mmHg for mean and 2.8 +/- 5.8 mmHg for diastolic measurements. The "limits of agreement" (mean difference +/- 2SD of bias) for both groups were within the range which would be acceptable for clinical setting. These results indicate that tonometry provides accurate, reliable and real-time blood pressure measurements together with continuous display of the pressure waveform for wide range of blood pressure. It was concluded that this system may replace the conventional invasive intra-arterial blood pressure monitoring system. This system is especially useful if it is combined with pulse oximetry and capnography.

摘要

目前已研发出一种基于动脉张力测量法的新型连续、无创动脉血压监测系统。该系统的传感器由15个压阻式压力传感器组成的阵列构成。自动传感器定位系统确保15个传感器中的至少一个位于动脉经气动压扁但未闭塞部分的中心位置。通过在70例血压正常的麻醉患者(正常血压组)和21例低血压麻醉患者(低血压组)中比较张力测量血压(TBP)与动脉内血压(IBP)测量值,对该系统的准确性、可靠性和临床可接受性进行了评估。通过Gould P23XL校准传感器在右侧或左侧桡动脉测量IBP,而在另一侧桡动脉测量TBP。两个系统的输出记录在PCM磁带录音机上以便后续分析。TBP波形与IBP波形相似。两种方法在收缩压、平均压和舒张压测量方面存在显著的高度相关性。正常血压组收缩压的偏差幅度(两种方法之间的平均差值)为-0.4±5.9(平均值±标准差)mmHg,平均压为-0.1±4.7 mmHg,舒张压为0.0±5.4 mmHg;低血压组收缩压的偏差幅度为-0.7±6.6 mmHg,平均压为0.7±5.0 mmHg,舒张压为2.8±5.8 mmHg。两组的“一致性界限”(偏差的平均差值±2倍标准差)均在临床环境可接受的范围内。这些结果表明,张力测量法能够提供准确、可靠且实时的血压测量,并能连续显示血压波形,适用于广泛的血压范围。得出的结论是,该系统可能会取代传统的有创动脉血压监测系统。如果该系统与脉搏血氧饱和度测定法和二氧化碳图相结合,将特别有用。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验